5
.Menu Interpretation
49
5.8.7 [Quasi-Peak]
Set the detector to the Quasi-Peak detector mode. This mode is
available when EMI filter is turned on. The quasi-peak detector is a
peak detector that is weighted by the duration and repetition rate of
the signal, as specified by the CISPR 16-1-1 standard. Quasi-peak
detection is characterized by a fast charge time and slow decay
time.
5.9
【
Display
】
hardkey
Key access:
Press
Shift key, then
Display
hardkey (
9
in
numeric keypad
).
Controls the screen display of the analyzer, such as full screen,
setting the on or off for window zoom, display line, amplitude scale,
grid and label.
5.9.1 [Full Screen]
Set to full-screen display graphical interface, press any key to exit.
5.9.2 [Zoom On Off]
In multi-window display mode, press this button to zoom in on the
selected window. Press the key for the first time to enlarge the
selected window to the entire graphic display area. Press this
button again to exit the entire graphic display area and restore the
multi-window display mode.
5.9.3 [UI Setting
►
]
Access the UI setting submenu.
5.9.3.1 [Display Line On Off]
When this menu is on, an adjustable horizontal reference line is