.Menu Interpretation
42
Activate the frequency counter function and display the count
results in the upper right corner of the screen. The counter counts
only the signals that are displayed on the screen.
1) [
Freq Count
On Off]
Turn on or off the frequency counter function. This function is
invalid when the tracking generator is activated. The count value is
displayed in the upper right corner of the screen.
2) [
BW
]
Sets the frequency counter resolution. The allowed values are 1 Hz,
10 Hz, 100 Hz and 1 kHz. Changing the counter resolution can
change the counter accuracy. The higher the resolution, the higher
the counting accuracy.
5.6.3
【
Marker→
►
】
Key access:
[Marker] bottom softkey, and then [Marker
→
►
] right softkey
Use the current marker readings to set the parameters (such as
Center frequency, Reference level).
5.6.3.1 [Mkr->CF]
Sets the center frequency of the analyzer to the frequency of the
current marker. This feature quickly moves the signal to the center
of the screen.
In Normal marker mode, the center frequency will be set to the
frequency of the current marker.
In Delta marker mode, the center frequency will be set to the
frequency at which the Delta Marker is located.
The function is not available in zero span mode.
5.6.3.2 [Mkr->CF Step]
Sets the center frequency step of the analyzer to the frequency of
the current marker.
In Normal marker mode, the center frequency step will be set to