4
.Quick Start
24
Pinch and spread vertically
to change the Scale/Div
4.12 Basic Measurement
Basic measurements include, input signal frequency and amplitude
display, marked by a frequency marker.
Follow these four simple steps below to implement input signal
measurement.
a)
Setting center frequency;
b)
Setting span and resolution bandwidth;
c)
Activate marker;
d)
Setting amplitude.
For example, to measure a 100 MHz, -10 dBm signal, you must
turn on the spectrum analyzer and ensure it is warmed up for 30
minutes to ensure measurement accuracy.
1. Equipment connection
Connect the output terminal of signal generator to the
RF Input
50Ω
connector on the top of the spectrum analyzer. Set the
parameters as follows:
Frequency
100 MHz
Amplitude -10 dBm
2. Setting parameters
1) Press
【
Preset
】
on the front panel to restore the analyzer to its