Power supply
Operation
0 °C/+32 °F to +40 °C/+104 °F
max. 90 % relative humidity, non-condensing
Input voltage
100 V~ to 240 V~
Mains frequency
50 Hz to 60 Hz
Output voltage
12 V
Battery charger
Reference number
4E60*
Storage and transport in original packaging
-25 °C to 70 °C/-13 °F to 158 °F
Storage and transport without packaging
-25 °C to 70 °C/-13 °F to 158 °F
Max. 93% relative humidity, non-condensing
Operation
5 °C to 40 °C/41 °F to 104 °F
Max. 93% relative humidity, non-condensing
Protection rating
IP40
Input voltage
12 V
Only when "DUAL" marking is not present (
see Page 66
)
Cockpit app
Article number
Cockpit 4X441-Andr=V*
Supported operating system
Android 4.0.3 or higher
Website for download
http://www.ottobock.com/cockpitapp
Data transfer
Wireless technology
Bluetooth 2.0
Range
approx. 10 m / 32.8 ft
Frequency range
2402 MHz to 2480 MHz
Modulation
GFSK
Data rate (over the air)
704 kbps
Maximum output power (EIRP)
-1.78 dBm (0.66 mW)
Only when "DUAL" marking is present (
see Page 66
Cockpit app
Reference number
4X441-IOS=* / 4X441-Andr=V* Cockpit
Supported operating system
iOS 9.3 / Android 4.1 or higher
Website for download
http://www.ottobock.com/cockpitapp
Data transfer
Wireless technology
Bluetooth Smart Ready
Range
approx. 10 m / 32.8 ft
Frequency range
2402 MHz to 2480 MHz
Modulation
GFSK, π/4 DQPSK, 8DPSK
Data rate (over the air)
2178 kbps (asymmetrical)
Maximum output power (EIRP):
+8.5 dBm
Torque values of the screw connections
Using a torque wrench, tighten the corresponding screws alternately in several cycles until the specified tightening
torque is reached.
Screw connection
Tightening torque
Tube adapter on prosthetic foot
15 Nm / 133 lbf. In.
Tube clamp of the knee joint
7 Nm / 62 lbf. In.
Proximal prosthesis components with pyramid receiver
15 Nm / 133 lbf. In.
Proximal prosthesis components with threaded connect
or
10 Nm / 89 lbf. In.
Flexion stop
0.6 Nm / 5 lbf. In.