Prior to delivery of the orthosis to the patient, the fit, support and all func
tions must be tested.
Pay attention to the parallel alignment of the orthosis joints.
Check the locking mechanism and instruct the patient in handling the lock
and the product.
INFORMATION
Instruct the patient not to unlock the orthotic knee joints under load.
6 Patient information
Inform the patient:
Verify lock functionality
►
Before standing up, check that both locks are engaged. Carry out a
function test.
Inspect the orthosis for damage
►
Inspect the orthosis for damage daily, paying special attention to the
locks and bars. Contact the qualified personnel promptly if changes are
noted.
7 Cleaning
System joints must be cleaned promptly after contact with water containing
salt, chlorine or soap, or when they get dirty.
1) Rinse system joints with clean fresh water.
2) Dry system joints with a cloth or allow to air dry. Avoid exposure to direct
heat (e.g. from an oven or radiator).
8 Maintenance
INFORMATION
The product may be exposed to increased loads by the patient.
►
Shorten the maintenance intervals according to the expected loads.
The manufacturer requires an inspection of the product every
6 months
to
verify functionality and check for wear at a minimum.
Only use 633F7 special lubricant.
A service kit can be ordered for product maintenance.
Spare parts are listed in the section "Components/design".
Before handing the orthosis over to the patient, secure all screw connec
tions with Loctite 241.
15
Summary of Contents for 17KL30 15-A
Page 2: ...1 2...
Page 103: ...3 2 3 3 36 3 4 4 4 1 4 2 103...
Page 104: ...150 C 300 C 104...
Page 105: ...10 C 60 C 5 743Y57 1 6 7 10 15 105...
Page 106: ...20 150 C 302 F 300 C 572 F 100 C 106...
Page 107: ...633F7 Loctite 241 6 7 1 107...
Page 108: ...2 8 6 633F7 Loctite 241 9 10 10 1 108...
Page 111: ...4 2 150 C 300 C 111...
Page 112: ...10 C 60 C 5 743Y57 1 6 7 10 15 20 mm 112...
Page 113: ...150 C 302 F 300 C 572 F 100 C 633F7 Loctite 241 6 113...
Page 114: ...7 1 2 8 6 633F7 Loctite 241 9 10 10 1 10 2 CE 93 42 EWG IX I VII 114...
Page 115: ...115...