No measured values in database
Configuration incorrect?
➝
Check configuration
1)
(for advanced operation, refer to online help).
Alkaline batteries (only OTT ecoLog 500): Battery voltage lower than
approx. 1.9 V
➝
Replace the batteries.
Lithium battery: Battery exhausted?
➝
Replace the battery.
Communication through cellular network not possible/
not successful
No transmission parameters (operator, etc) configured
➝
Check the operating parameters entered, refer to Chapter 9.4.
A remote communication connection cannot be established with the
OTT ecoLog 500 3G
➝
this is not a malfunction. In a 3G cellular network (UMTS/HSPA+), this is no
longer possible for technical reasons. The circuit-switched service required for
this purpose, i.e. CSD (Circuit Switched Data) is no longer available in 3G cel-
lular networks.
Alkaline batteries (only OTT ecoLog 500): Battery voltage lower than
approx. 1.9 V
➝
Replace the batteries.
Lithium battery: Battery exhausted?
➝
Replace the batteries.
SIM card missing
➝
Insert the SIM card, refer to Chapter 6.
Incorrect PIN
➝
Check PIN, refer to Chapter 9.4.
GSM signal too low
➝
Check whether an external radio antenna should be used, refer to
Chapter 7.
No measured values present (by transmitting data)
➝
Wait for the next data interval.
1)
The default factory configuration is located on the "OTT Water Logger Software" CD-ROM.
47