Other displays in the "Basic operation" window
Measuring range Measuring range of the pressure probe, refer to Fig. 1.
System length Cable length including communication unit/pressure probe, refer to Fig. 1.
Date/time Internal date/time of the OTT ecoLog 500 (3G).
Battery lifetime (button) The operating program calculates the approximate life of the different battery
types based on the settings made (at –20 and +20 °C ambient temperature in
each case).
Detailed information on the "Advanced operation" function can be found in the online help.
Factory settings
OTT ecoLog 500 (3G)
Site number OTT ecoLog 500 (3G) serial number
Site name ecoLog 500 1
Water level/Pressure Temperature
Parameter number 0001 0002
Measurement type Water level · depth
Unit m °C
Decimal places 2 1
Sample interval 1 h 1 h
Storage interval 1 h 1 h
Modem connected Enabled
Authorize roaming Disabled
SMS data transmission active Disabled
GPRS(/3G) data transfer active Disabled
Store GSM signal strength Disabled
In addition, the OTT ecoLog 500 (3G) records the supply voltage at a sample and storage interval of 1 hour
(refer to online help).