OPERATION SETTINGS
41
OI-6000K-X-X-X-NXP-X OPS_GUIDE_REV 1.0
5.
Affix a regulator to the calibration gas bottle.
6.
Attach the tubing on the Calibration Adapter Kit to the regulator on the calibration gas bottle.
7.
Ensure that the gas is flowing and press the
MENU
button to begin calibrating the sensor, the
unit will automatically begin the calibration process, the amount of time on the timer will vary
based on the gas type. During calibration, the display will show a countdown of the time
remaining until the process is complete.
NOTICE
Once the calibration countdown has started, the process cannot be stopped without
disconnecting the power from the unit.
8.
When calibration is complete, detach the Calibration Adapter Kit from the sensor housing and
reaffix the sensor housing cap. Press the
MENU
button to advance to the Low Alarm Setting
screen.