INSTALLATION AND STARTUP
9
OI-6000K-X-X-X-NXP-X OPS_GUIDE_REV 1.0
CUSTOMER POWER
SUPPLY WIRING
OI-6000K
BACKPLANE TERMINAL BLOCK
+12 to +35 VDC
RED
+12 to +35 VDC
RED
GND
BLACK
GND
BLACK
4-20 mA
4-20 mA
NOTICE
Wiring power to the device is the
ONLY
requirement for the OI-6000K to operate. With the
provision of power, the unit will function normally, indicating the presence of toxic gas at the
sensor and providing the gas level reading on the display screen. To utilize the added
functionality of the device, additional wiring is necessary. If an Otis Monitor is not used, the OI-
6000K can be powered from any +12 to +35 VDC power supply that can supply at least 1.250
Amps.