6
of the AFO.
6. To relieve contact between the medial strut and the apex of the medial
malleolus, grind the edge of the strut according to the shape shown (figure 8).
- you may grind up to 1 cm into the strut
- stay at least 7 cm from the footplate
GENErAL rEMArKS AND ADDItIONAL ADVICE:
1. To shorten the footplate, cut or grind off the appropriate length. Be certain to
buff the edges to eliminate sharp edges. To lengthen the
footplate, use foot insert of choice, and fixate it to the foot plate.
2. To adjust the biomechanics of the foot, the use of the correct insert is advised.
3.
Contraindications:
Do not use the AFO in case of excessive ankle instability
or in case excessive deformities to the foot where pressure relief cannot be
solved by additional padding. It is contraindicated to use the AFO in case of
fluctuation oedema and in case of severe spasticity.
4. The AFO cannot be reshaped by heating. Heating will cause delaminating and
negatively alter the dynamics of the orthosis.
5. The right footwear is critically important for proper function of the AFO. The
following criteria should be met when selecting the right
footwear to use with this AFO:
• Firm heel counter - for proper control of the rear foot
• Firm shank - to take stress off the AFO
• Laced or Velcro fastening - for easier donning and doffing, and to allow
adjustable compressive support over the mid-foot
• The shoes have to be wide enough to allow proper fitting of the AFO
inside the shoe
• Removable insole - to allow space for the AFO footplate with
modifications
• Approximately 6/9“ (1,7 cm) heel height - as a starting point to control
the knee extension
For optimal comfort and function as well as proper customization of the ÖSSUR
AFO: Dynamic, the use of an orthopaedic insole is recommended.
wArNING
With the use of this type of device, it is possible to receive various types of skin
irritation and or pressure points. If you experience any unusual discomfort while
using this product, consult your health care provider immediately.
This product has been designed and tested based on single patient usage and is
not recommended for multiple patient use. If any problems occur with the use of
this product, for example, if unusual movement or product wear is detected in a
structural part of the product, immediately discontinue use of the device and
contact your medical professional. Customization beyond that specified in this IFU
is not recommended.
Summary of Contents for AFO Dynamic
Page 1: ...I n s t r u c t i o n s f o r u s e AFO Dynamic...
Page 4: ...3 1 cm 7 cm 1 2 3 4 5 6 8 7 A B B C D E...
Page 31: ...30 1 2 3 AFO AFO 4 AFO 5 AFO AFO AFO AFO AFO 1 7 cm SSUR AFO Dynamic 1 IFU...
Page 33: ...32 6 8 1 7 1 2 3 AFO AFO 4 AFO 5 AFO AFO AFO Velcro AFO AFO 6 9 1 7 SSUR AFO...
Page 34: ...33 A B C D E AFO 1 AP AFO AFO AFO 3 AFO 90 AFO 4 C P O 2 2 5 6 AFO 1...
Page 36: ...35...