8. INSTALLATION
Overview
This section provides detailed installation instructions for the
CLOVER A1c
®
.
You must follow the installation steps correctly to ensure proper installation, operation,
and service.
Place the system where it will not be subjected to extreme temperature variations. Avoid
proximity to open windows, direct sunlight, ovens, hot plates, open burners, radiators
and dry ice baths.
Do not place the
CLOVER A1c
®
on the same bench with a source of vibration.
Provide bench space large enough to allow free air circulation around the system - 8cm
(3 inches) on all sides.
Your
CLOVER A1c
®
is delivered in 1 shipping carton.
1. Carefully remove the contents of the shipping carton.
2. Inspect the carton and system for visible signs of damage.
3. If damage to the system exists, immediately file a complaint with the carrier.
4. Make sure all items are included with your system, and keep them for future use.
CAUTION
Do not drop or handle the system roughly. This can disturb internal calibrated optics
and electronics or cause other damages.
Always handle the system with care. The
CLOVER A1c
®
Self is a precision system
and must be handled accordingly.
Connect the System Power
1. Ensure that the system power button is in the off position.
2.Connect the power code of the system to appropiately grounded AC electrical outlet.
Installing Connections
1
2
3
4
5
System Connections:
1. Power Button
2. DC 9V adapter port
3. USB port (PC connection port)
4. Barcode Scanner port
5. Thermal Printer port (Serial port RS 232)
9. OPERATION
9.1 Power On
Connect the DC 9V adapter to the power port on the back side of the analyzer.
Make sure that the lid is closed before connecting the adapter.
On/off 9V
USB Barcode Printer
After the system is properly installed, you can start the
CLOVER A1c
®
.
To start the
CLOVER A1c
®
, turn the power
switch to the 'ON' position.
If the lid is open, the icon of "Close the lid "
will be displayed.
Close the lid, and the warming up will start.
If the cartridge is inserted, the icon of
"Remove the cartridge" will be displayed.
Remove the cartridge and close the lid.
Warming up will start.