13. CHECKING THE SYSTEM
QUALITY CONTROL
The
CLOVER A1c
®
Check Cartridge checks that the optical and operating systems of the
analyzer.
Type of Check Cartridges
•
CLOVER A1c
®
Daily Check Cartridges
•
CLOVER A1c
®
Monthly Check Cartridges
Storage Instruction
• The Check Cartridge must be protected from sunlight during storage.
• Store the Check Cartridge at temperature 2-32
ఁ
(36-90
ሥ
).
• Store the Check Cartridge at humidity < 90%
.
• Always store the Check Cartridge in its protective packaging to prevent scratches which
may affect the result.
• If refrigerated allow Check Cartridge reach room temperature for at least 1 hour before
use.
Precautions/ Warnings
• For In Vitro Diagnostic use.
• Do not use the Check Cartridge beyond the expiration date printed on the Check
Cartridge Label and Pouch.
• Do not use the Check Cartridge if stored incorrectly or if it is dirty, scratched or damaged.
13. CHECKING THE SYSTEM
13.1 Daily Check Cartridge
When to Use the Daily Check Cartridge
• Once a day before samples are tested.
• After the analyzer has been moved.
• After an error message. (Er 1 or Er 3)
The Daily Check Cartridge, used to check the analyzer, is composed of a cartridge
in a pouch, without a reagent pack.
Life-time : Up to the expiry date on the cartridge label.
NOTE
:
Be careful in maintaining the Daily Check Cartridge by keeping in the
provided blue color pouch when not in use.
*
The Daily Check Cartridge can be purchased from local distributors
。