15. TROUBLESHOOTING
Unplug the power adaptor and reconnect.
If the problem still persists,
contact your local representative.
There is a problem
with the power
adaptor or the
analyzer itself
The
Analyzer
does not
power on
>14%
Please use the check cartridge to
validate the analyzer and re-test the
blood sample.
The HbA1c test
result is higher than
14.0%
(130 mmol/mol)
ERROR CODE
CORRECTIVE ACTION
DESCRIPTION
Do not
Reuse
Problem with reusing
the cartridge.
(Test or Monthly
Check Cartridge)
Please press mode button and remove the
cartridge.
Check whether the cartridge is reused or not.
Insert a new cartridge and perform the test again.
If the problem persists contact your local
representative.
16. SPECIFICATION
Storage Temperature
10 ~ 35 (50 ~ 95 )
Option
PC Connection Cable
Thermal Printer
Barcode Scanner
Relative Humidity Range
10% ~ 90%
Operating Temperature
17 ~ 32 (63 ~ 90 )
Dimensions/Weight
200*198*139(mm) / 1.4kg
Power Required
DC 9 V-2 A
Wavelength
415 nm (normal)
Memory Capacity
200 test results
Reading Time
5 minutes
Test Range
4.0~14.0%
Sample Volume
4 uL
S a m p l e T y p e
Capillary whole blood,
Venous blood with anticoagulant