13. CHECKING THE SYSTEM
CAUTION :
5) Insert the Check Reagent Pack, when the ‘Insert Check Reagent
Pack’ icon is displayed, Insert it respectively.
4) Gently mix the Check Reagent Pack before use.
Do not mix too vigorously, it may cause air bubbles.
If bubbles are present wait until they disappear.
6) Close the lid. The test starts automatically.
13/03/30
13/03/30
13/03/30
13. CHECKING THE SYSTEM
7) After 5 minutes ‘OK’ or error message will be displayed.
8) Press ( ) and remove the Monthly Check Cartridge when the
test is completed.
NOTE :
If an error message is displayed, repeat the test again. If it still
displays error message, do not use the analyzer. Please contact
your local representative for customer support.