3.3
Detect Mode
radEAGLE
User Manual
Settings and Commands
Reset
Reset the timer and clear the spectrum recorded so far.
Analyze
Initiate the analysis of the spectrum before the preset recording duration elapsed.
Continue
Continue with the Easy ID spectrum acquisition after an analysis. Sometimes, the
preset duration was simply too short or the source is very weak, so extending the
duration gives additional trust in the result.
Exit
Return to the superior screen.
3.3 Detect Mode
The Detect measuring mode is a tool to locate radiation sources by giving rapid visual and
audio feedback to the changing dose rate of incoming radiation.
The tool starts with measuring the background radiation (Fig.
, p.
Figure 26:
Detect mode background measurement
It is preferable to take the background in a low radiation environment. If you are in a
facility with a high natural background or with multiple radiation sources present, the
detection capability adapts to this situation.
After finishing the background acquisition, the display shows a chart with colored bars rep-
resenting the count rate history of the last couple of seconds (Fig.
, p.
).
Green Bars
represent radiation levels close to the background.
Red Bars
represent increased radiation potentially caused by a source (Fig.
, p.
34/147
Software 3.2.12 • Document 3.3.0o
© innoRIID GmbH • 2019-04-11
Summary of Contents for radEAGLE
Page 1: ...radEAGLE User Manual Software 3 2 12 Document 3 3 0o 2019 04 11 ...
Page 20: ...20 147 Software 3 2 12 Document 3 3 0o innoRIID GmbH 2019 04 11 ...
Page 40: ...40 147 Software 3 2 12 Document 3 3 0o innoRIID GmbH 2019 04 11 ...
Page 106: ...106 147 Software 3 2 12 Document 3 3 0o innoRIID GmbH 2019 04 11 ...
Page 118: ...118 147 Software 3 2 12 Document 3 3 0o innoRIID GmbH 2019 04 11 ...
Page 134: ...134 147 Software 3 2 12 Document 3 3 0o innoRIID GmbH 2019 04 11 ...