E.4
Unmanned Neutron Detection Testing
radEAGLE
User Manual
E.4
Unmanned Neutron Detection Testing
This item is available for
radEAGLE
models with a neutron detector (see
, p.
For tests involving an un-moderated neutron source, an appropriate phantom of Polymethyl
Methacrylate (PMMA, (C
5
H
8
O
2
)
n
, acrylic glass) or equivalent must be placed between the
neutron source and the instrument to accurately simulate the moderation effects of field
mission environments (which always provide moderation through surrounding material).