Moxi Flow
™
User Guide
Page 2
Safety Precautions
The Moxi Flow
TM
has integrated safety features that are designed for the protection of the user.
Use the Moxi Flow
TM
only as directed in this guide. Please review and understand the safety
instructions below before operating the system.
General Safety
WARNING
To avoid the danger of electric shock:
•
Prior to use, verify that the USB cable and USB charging adapter are plugged
securely into a properly grounded AC power outlet. Verify that the connection
between the USB cable and the instrument is secure and the AC power in your
location is within the specifications for the instrument (see page 31).
•
Do not immerse the instrument, USB cable, or USB power adapter in liquid or allow
liquid to enter the instrument.
•
Do not attempt to disassemble or service the Moxi Flow
TM
. The instrument has no
user serviceable parts. All service must be performed by ORFLO Technologies.
WARNING
To avoid the danger of fire or explosion, do not use the Moxi Flow
TM
with
flammable or explosive liquids.
Laser Safety
WARNING
The Moxi Flow
TM
is a Class 1 laser product in accordance with IEC 60825-
1:2014. The system contains a Class IIIb laser that operates at 532 nm with
a maximum output power of 50 mW. Direct exposure to laser radiation is
harmful. The following guidelines should be followed to avoid exposure:
•
An interlock on the upper door prevents the laser from operating when the door is
open. Do not attempt to defeat the interlock and run a test with the door open.
•
Do not operate the Moxi Flow
TM
if there is any damage to the instrument’s enclosure.
•
The laser in the instrument is not adjustable or serviceable. Attempts to adjust or
service the laser may result in hazardous radiation exposure.