(blue) gates is displayed in the black box on the bottom left (e.g. “Live Cells”). Information
pertinent to the population below the fluorescent (red) gate and between the size (blue) gates is
displayed in the black box on the bottom right of the scatter plot (e.g. “Dead Cells”). Total
particle counts in the size-gated region are displayed above the scatter plot.
Store and Recall of Gate Settings
By default, the Moxi Flow™ gates are “sticky” meaning that each new test defaults to the gate
settings (size and fluorescence) of the prior test. The Moxi Flow™ also allows the user to store
up to five gate locations in memory for recall. A given set of gate settings can be saved by
hitting the
STO
button. Doing so brings up a list of five memory locations (numbered zero
through four, see images below). Memory locations that are already occupied are shown in red,
empty locations are shown in black. Touching either red or black locations will result in that
location being overwritten by the current fluorescent and size gate settings.
Gates can later be recalled at any point (after a test or for a saved test) by selecting the
RCL
button. Available memory bank locations are listed in green (see image below). Selecting one
of the buttons will automatically apply the gate parameters that were saved at that current
location.