ERROR STATEMENTS
9.
ERROR
CAUSE
ELIMINATION
E. D. UN.
Number is too small (large negative) to be displayed
change DP setting, channel constant setting
E D. OV.
Number is too large to be displayed
change DP setting, channel constant setting
E. T. UN.
Number is outside the table range
increase table values, change input setting (channel
constant setting)
E. T. OV.
Number is outside the table range
increase table values, change input setting (channel
constant setting)
E. I. UN.
Input quantity is larger than permitted input quantity
range
change input signal value or input (range) setting
E. I. OV.
Input quantity is larger than permitted input quantity
range
change input signal value or input (range) setting
E. HW.
A part of the instrument does not work properly
send the instrument for repair
E. EE
Data in EEPROM corrupted
perform restoration of manufacture setting, upon repeated
error statement send instrument for repair
E.. SET.
Data in EEPROM outside the range
perform restoration of manufacture setting, upon repeated
error statement send instrument for repair
E. CLR.
Memory was empty (presetting carried out)
upon repeated error statement send instrument for repair,
possible failure in calibration
E.. OUT.
Analogue output current loop disconnected
check wire connection