Electrical Installation
65
-
The industrial process interface provides the following inputs and outputs:
Name
Description
max range
1)
/ status
A IN 1 / 2
Analog input 1 and 2
0-10 V
2)
D IN 1
Digital input
(active-
low = 0…0,6 V)
24 V
AO1 / 2 / 3
Analog output 1, 2 and 3
Alarm output 1, 2 and 3
0/4-20 mA
DO1 / 2/ 3
Relay output 1, 2 and 3
3)
open/ closed (red LED on) / 0...30 V, 400 mA
FS
Fail-safe relay
open/ closed (green LED on)/ 0...30 V, 400 mA
1)
depending on supply voltage; for 0-20 mA on the AO the PIF has to be powered with min. 5V < (1.5 + working resistance * 0.021) <
24 V; Example: R
Load
= 500 ohm
→ U
min
= 1.5 + 500 * 0.021 = 12 V, R
Load
= 100 ohm
→ U
min
= 1.5 + 100 * 0.021 = 3.6 V
→ min. 5 V
2)
the AI is designed for max. 24 V, the voltage level above 10 V is not interpreted
3)
active if AO1, 2 or 3 is/ are programmed as alarm output
The alarm output can be configured as a threshold between
0-4 mA
for
no alarm
and between
10-20 mA
as
alarm
. For values outside the respective range, the relay does not switch on the
DO.
Summary of Contents for Xi 400
Page 1: ...Operator s Manual optris Xi 80 400 410 Spot finder IR camera...
Page 27: ...Mechanical Installation 27 Figure 5 Xi 80 dimensions mm...
Page 28: ...28 Figure 6 Xi 400 dimensions mm...
Page 29: ...Mechanical Installation 29 Figure 7 Xi 410 dimensions mm...
Page 34: ...34 Figure 11 Mounting bracket ACXIAPLAB dimensions mm weight 276 g...
Page 36: ...36 Figure 12 Water cooling ACXIW and mounting kit ACXIxxxWAK1 dimensions mm weight 1710 g...
Page 40: ...40 Figure 15 Shutter ACXIS dimensions mm weight 550 g...
Page 71: ...Functions 71 Figure 35 Possibilities of power supply for Xi 80 410 via Ethernet connection...
Page 79: ...Functions 79 Figure 37 Configurations menu Device PIF Figure 38 Setup...
Page 92: ...92 7 2 Software window Figure 45 Software window 1 2 3 2 5 6 7 9 4 10 11 12 1 8...
Page 107: ...Appendix A Table of emissivity for metals 107 Appendix A Table of emissivity for metals...
Page 108: ...108...
Page 118: ...118 Figure 58 Relay output at industrial PIF...
Page 119: ...Appendix G Declaration of Conformity 119 Appendix G Declaration of Conformity...
Page 120: ...optris Xi MA E2021 08 A...