Emissivity
103
-
Figure 51:
Spectral emissivity of several materials:
1
Enamel,
2
Plaster,
3
Concrete,
4
Chamotte
If the emissivity chosen is too high, the infrared thermometer may display a temperature value which is much
lower than the real temperature
– assuming the measuring object is warmer than its surroundings. A low
emissivity (reflective surfaces) carries the risk of inaccurate measuring results by interfering infrared radiation
emitted by background objects (flames, heating systems, chamottes). To minimize measuring errors in such
cases, the handling should be performed very carefully, and the unit should be protected against reflecting
radiation sources.
Summary of Contents for Xi 400
Page 1: ...Operator s Manual optris Xi 80 400 410 Spot finder IR camera...
Page 27: ...Mechanical Installation 27 Figure 5 Xi 80 dimensions mm...
Page 28: ...28 Figure 6 Xi 400 dimensions mm...
Page 29: ...Mechanical Installation 29 Figure 7 Xi 410 dimensions mm...
Page 34: ...34 Figure 11 Mounting bracket ACXIAPLAB dimensions mm weight 276 g...
Page 36: ...36 Figure 12 Water cooling ACXIW and mounting kit ACXIxxxWAK1 dimensions mm weight 1710 g...
Page 40: ...40 Figure 15 Shutter ACXIS dimensions mm weight 550 g...
Page 71: ...Functions 71 Figure 35 Possibilities of power supply for Xi 80 410 via Ethernet connection...
Page 79: ...Functions 79 Figure 37 Configurations menu Device PIF Figure 38 Setup...
Page 92: ...92 7 2 Software window Figure 45 Software window 1 2 3 2 5 6 7 9 4 10 11 12 1 8...
Page 107: ...Appendix A Table of emissivity for metals 107 Appendix A Table of emissivity for metals...
Page 108: ...108...
Page 118: ...118 Figure 58 Relay output at industrial PIF...
Page 119: ...Appendix G Declaration of Conformity 119 Appendix G Declaration of Conformity...
Page 120: ...optris Xi MA E2021 08 A...