112
-
RS485 for Xi 80 and Xi 410
The Xi 80 and Xi 410 cameras support a direct RS485 connection. RS485 interface is designed for high-
speed serial data transmission over long distances and as a bidirectional bus system with up to 32
participants. An RS485 bus can be set up as a 2-wire as well as a 4-wire system. For the Xi 80/410 cameras,
it is designed for a 2-wire system. A maximum cable length of 500 meters can be realized. At least a master
and slave is necessary for communication. The master can be a PLC system and the slave is the camera.
RS485 is supported for Xi 80 cameras which have the hardware version 3001, firmware 3023 and at least
the serial number which starts with 1812xxxx. For the Xi 410 it is working for all cameras starting at firmware
version 3815. Here it is important that the communication is just working via Ethernet connection or in
autonomous mode, but not with USB.
The connection for RS485 can be done directly on the terminal block (A and B) which is provided in the
scope of supply of the camera.
Command list
The command list and more information can be found on the supplied USB stick under
Documentation and Manuals (Serial Communication Description-RS485-xxxx-xx-x). Every
command must expire with CR/LF (0x0D, 0x0A).
Summary of Contents for Xi 400
Page 1: ...Operator s Manual optris Xi 80 400 410 Spot finder IR camera...
Page 27: ...Mechanical Installation 27 Figure 5 Xi 80 dimensions mm...
Page 28: ...28 Figure 6 Xi 400 dimensions mm...
Page 29: ...Mechanical Installation 29 Figure 7 Xi 410 dimensions mm...
Page 34: ...34 Figure 11 Mounting bracket ACXIAPLAB dimensions mm weight 276 g...
Page 36: ...36 Figure 12 Water cooling ACXIW and mounting kit ACXIxxxWAK1 dimensions mm weight 1710 g...
Page 40: ...40 Figure 15 Shutter ACXIS dimensions mm weight 550 g...
Page 71: ...Functions 71 Figure 35 Possibilities of power supply for Xi 80 410 via Ethernet connection...
Page 79: ...Functions 79 Figure 37 Configurations menu Device PIF Figure 38 Setup...
Page 92: ...92 7 2 Software window Figure 45 Software window 1 2 3 2 5 6 7 9 4 10 11 12 1 8...
Page 107: ...Appendix A Table of emissivity for metals 107 Appendix A Table of emissivity for metals...
Page 108: ...108...
Page 118: ...118 Figure 58 Relay output at industrial PIF...
Page 119: ...Appendix G Declaration of Conformity 119 Appendix G Declaration of Conformity...
Page 120: ...optris Xi MA E2021 08 A...