Mechanical Installation
31
-
3.2 Accessories
3.2.1
Air purge laminar
The lens must be kept clean at all times from dust, smoke, fumes and other contaminants in order to avoid
reading errors. These effects can be reduced by using an air purge. (
Part-No.: ACXIAPL
)
•
Make sure to use oil-free, technically clean air, only.
•
The needed amount of air (approx. 2...10 l/min.) depends on the application and the
installation conditions on-site.
•
The laminar air purge has a Si-protective window. Typical transmission value: 0.82
(deviations possible), replacement window available under the
Part-No.: ACXIAPLPWSI
•
The corresponding mounting bracket (
Part-No.: ACXIAPLAB
) is mandatory.
•
Material: Anodized aluminum, weight: 218 g / 494 g with mounting bracket
•
Ambient temperature: 0...80 °C (T
Amb
camera: 0...50 °C); with water cooled housing up to
250 °C
Air flow
The air purge can be mounted in
four different positions.
The direction of the airflow must
always be clear.
Summary of Contents for Xi 400
Page 1: ...Operator s Manual optris Xi 80 400 410 Spot finder IR camera...
Page 27: ...Mechanical Installation 27 Figure 5 Xi 80 dimensions mm...
Page 28: ...28 Figure 6 Xi 400 dimensions mm...
Page 29: ...Mechanical Installation 29 Figure 7 Xi 410 dimensions mm...
Page 34: ...34 Figure 11 Mounting bracket ACXIAPLAB dimensions mm weight 276 g...
Page 36: ...36 Figure 12 Water cooling ACXIW and mounting kit ACXIxxxWAK1 dimensions mm weight 1710 g...
Page 40: ...40 Figure 15 Shutter ACXIS dimensions mm weight 550 g...
Page 71: ...Functions 71 Figure 35 Possibilities of power supply for Xi 80 410 via Ethernet connection...
Page 79: ...Functions 79 Figure 37 Configurations menu Device PIF Figure 38 Setup...
Page 92: ...92 7 2 Software window Figure 45 Software window 1 2 3 2 5 6 7 9 4 10 11 12 1 8...
Page 107: ...Appendix A Table of emissivity for metals 107 Appendix A Table of emissivity for metals...
Page 108: ...108...
Page 118: ...118 Figure 58 Relay output at industrial PIF...
Page 119: ...Appendix G Declaration of Conformity 119 Appendix G Declaration of Conformity...
Page 120: ...optris Xi MA E2021 08 A...