  c) Keep the microscope away from dust. When it is not used, please put the dust cover 

over it.

   d) Keep the microscope away from vibration.
   2) Insert the micro-slide specimen into the clamp(5). Be sure that the cover glass is 
facing towards the objective. Otherwise you will not be able to focus your specimen at 

using higher powered objective(40X, or 100X).

   3) Place a lower powered objective(4X, or 10X) into position and simply turn on the 

illuminator.

Caution: 

   The power supply voltage must be fitted to the microscope; Otherwise it will 

damage the circuit and bulb, even lead to insecurity.

   4) Rotate the knobs(14,15) to move the specimen so that it is centred over the in-stage 

condenser.

   5) Focus the objective on the specimen by turning the coarse focusing knob(24) until the 

image of your specimen is bright and clear. You can find the focal plane and focusing 
upwards by using the lower powered objective, and then you can bring the specimen into 

sharp focus by turning the fine focusing knob(23).

   6) The specimen now is in sharp focus. Rotate the nosepiece(1) to the other objectives 

and focus using only the fine focusing knob(23). Since the optics on the microscopes are 

Par-focal and Par-centered, only slightly turn the fine focusing knob(23) to make the image 
bright and clear. 

Note:

    It is important to note that the 4X and 10X objectives can never come into 

contacting with your micro-slide specimen because of our built-in stop. The 40X 
and 100X objectives may occasionally touch the micro-slide specimen. But because 

they have retractable mounts, the micro-slide specimen will not be damaged.

   7) To make the image clearer, you can adjust diameters of the iris diaphragm to match 
with the numerical aperture of objective in use(Sction 8. ABBE condenser in P.5).

 

 

E) Trouble Shooting:

   If there are some troubles in oberating, recheck the instrument carefully as the following 

describing before connect us or our representative in you area.

   1) Troubles in operating:

Can't focus in using high powered 
objective.

The specimen is mounted on the 
stage upside down or the cover 
glass is too thick.

Reverse the specimen or choose the 

standard cover glass(0.17mm).

The objective always touches with 
the slide when changing.

The cover glass is too thick.

Choose the standard cover glass 
(0.17mm).

Move the specimen no smoothly.

The clamp is not fixed stably.

Fix the clamp stably on the stage.

Interpupillary distance is not 

correctly adjusted.

Dioper adjusting is incomplete.

The brightness is not suitable.

Checkthe illuminator and adjust its 
brightness.

The specimen goes out of focus.

Adjust the the upper focusing limit.

Troubles

Causes

Remedies

The stage is limitted too low.

The slide is often broken by 
objective

P.7

Adjust the the upper focusing limit.

The stage is limitted too high.

Incomplete binocular vision.

Correct the interpupillary distance .

Complete the dioper adjusting.

Summary of Contents for A11.1303 Series

Page 1: ...www optoecu com 1 15 sale cnoec com A11 1303 Student Biological Microscope Instruction Manual...

Page 2: ...oubleshoot them by yourself according the manual please contact us or our representative in you area 3 Safety Before change a bulb or need to open the base ensure that the microscope has been disconne...

Page 3: ...Focusing System P 5 10 Illumination System P 5 D How to Use and Assemble P 6 E Trouble Shooting P 7 1 Troubles in Operating P 7 2 Troubles in Optical System P 8 3 Troubles in Electric System P 8 F Mai...

Page 4: ...F20X or Huygenian eyepieces as 5X 6X 10X 12 5X 15X 16X The specifications of WF10X are showed as following P 1 a b e c d f Fig 1 The Principle Draft In this microscope the standard outfit of the objec...

Page 5: ...Down 17 Swtich 18 Plate for Adjusting Brightness 16 Screw for Bulb Changing 8 Screw for Centring the Condenser Fig 2 The Structure Draft 3 Mechanical tube length 160mm 4 Conjugated distance between o...

Page 6: ...process software help you to preview measure compare count save and delete Connecting the CCD camera with the TV the micro image will be showed for teaching V type dual observing head 360 rotatable On...

Page 7: ...screen User friendly sophisticated micro image process software help you to preview measure compare count save and delete 3 Insert the CCD camera with the CCD Linker Connecting the CCD camera with th...

Page 8: ...the knob 25 built in rack and pinion mechanical system controls the condenser up or down Usually raise the condenser higher when 100X or 40X objective is used Down the condenser lower when 10X or 4X...

Page 9: ...b D How to use and assemble 1 Unpack the microscope and its parts carefully Check and sort out all parts according to the packing list 2 For the convenience of packaging and delivering the components...

Page 10: ...e 4X and 10X objectives can never come into contacting with your micro slide specimen because of our built in stop The 40X and 100X objectives may occasionally touch the micro slide specimen But becau...

Page 11: ...ld is dark Image moves while focusing The specimen is not caught stably by the clamp The image is yellow Blue filter is not used The viewing field is too dark P 8 Remove the dust and dirt There are du...

Page 12: ...to choose model XSP21 01V or XSP21 01T microscope 1 See Fig 4 and connect the CCD camera with the microscope as following steps 1 Take the plastic dust cover off from the viewing head and screw the v...

Page 13: ...nd its PL pilot lamp is lighting 3 Whether the TV is turned on the mode AV or VIDEO 4 Whether the rod on the right of trinocular head is pushed out 5 Whether the microscope is focused correctly At thi...

Page 14: ...can observe by using common eyepiece at the same time of showing the micro image in the PC screen The electron eyepiece is also used in model XSP21 01M or XSP21 01B microscope See Fig 6 and connect th...

Page 15: ...een as the following ways 1 Turn the plate 19 to make the illuminator darker 2 Turn the plate 10 to make the diameter of the iris diaphragm smaller 3 Rotate the knob 23 to lower the ABBE condenser Whe...

Reviews: