Stereomicroscopes

Stemi DV4

Stemi DV4 SPOT

Stemi DR 1040
Stemi DR 1663

Operating Instructions

Summary of Contents for Stemi DR 1040

Page 1: ...Stereomicroscopes Stemi DV4 Stemi DV4 SPOT Stemi DR 1040 Stemi DR 1663 Operating Instructions...

Page 2: ...s not covered by an update service Unless expressly authorized forwarding and duplication of this document and the utilization and communication of its contents are not permitted Violations will entai...

Page 3: ...2 2 4 Technical Data 2 4 2 3 Modular Zeiss Stemi system 2 6 2 3 1 System diagram extract 2 6 2 3 2 Interfaces in the modular system 2 7 3 OPERATION 3 1 3 1 Basic instrument as compact version with st...

Page 4: ...LR camera 3 12 3 2 3 2 SONY camera Digital Handycam DCR PC100 3 13 3 2 3 3 SONY camera Digital Still Camera Cybershot DSC F505 3 14 3 2 3 4 SONY camera Digital Still Camera Cybershot DSC S70 3 15 3 2...

Page 5: ...ected light 3 7 Fig 3 6 Changing the lamp reflected light 3 7 Fig 3 7 Removing the glass plate 3 8 Fig 3 8 Changing the lamp transmitted light 3 8 Fig 3 9 Inserting the glass plate 3 9 Fig 3 10 Settin...

Page 6: ...ILLUSTRATIONS Carl Zeiss Stemi DV4 DR 0 6 B 40 003 e 08 00...

Page 7: ...ppendix I of directive 73 23 EC The instruments meet the requirements of the EC directive 89 336 EC and the EMC legislation of November 9th 1992 This operation manual includes information and warnings...

Page 8: ...3 Notes on warranty The manufacturer guarantees that the instrument has no material and production defects when delivered You must inform us of any defects immediately and we must do anything to mini...

Page 9: ...onverted and upgraded in accordance with the modular system of Zeiss stereomicroscopes Interchangeable optics components permit the magnification range to be varied as required Modern illumination equ...

Page 10: ...INTRODUCTION Carl Zeiss Stemi DV4 DR 1 4 B 40 003 e 08 00...

Page 11: ...with stand C However this is barely useful since stand C is already equipped with reflected light illumination The four microscope bodies differ in the following features Stemi DV4 Continuous magnific...

Page 12: ...DESCRIPTION Carl Zeiss Compact version with stand C Stemi DV4 DR 2 2 B 40 003 e 08 00 2 2 Compact version with stand C 2 2 1 System diagram Fig 2 2 System diagram of compact version with stand C...

Page 13: ...r reticles dia 19 mm Mount dia 76 mm for microscope body Mount dia 53 mm for accessories Internal thread M50x0 75 for mount for attachment systems dust cover or annular light adapter Mount dia 84 mm f...

Page 14: ...lative humidity 20 to 85 at 35 C Atmospheric pressure 800 hPa to 1060 hPa Operation Permissible ambient temperature 10 to 40 C Permissible relative humidity maximal 85 Atmospheric pressure 800 hPa to...

Page 15: ...12 0x 20 0x 6 3x 25 2x 10 0x 40 0x 12 5x 50 0x 16 0x 64 0x 20 0x 80 0x Object field in mm 66 7 16 7 50 0 12 5 66 7 40 0 16 7 10 0 31 8 7 9 20 0 5 0 16 0 4 0 12 5 3 1 10 0 2 5 Stemi DR 1663 Eyepiece W...

Page 16: ...ON Carl Zeiss Zeiss Stemi modular system Stemi DV4 DR 2 6 B 40 003 e 08 00 2 3 Modular Zeiss Stemi system 2 3 1 System diagram extract Fig 2 4 Stemi DV4 DR system diagram within the Zeiss Stemi modula...

Page 17: ...les Mount dia 76 mm for microscope body Mount dia 53 mm for accessories Internal thread M50x0 75 for mount for attachment systems dust cover or annular light adapter Mount dia 84 mm for stages insert...

Page 18: ...DESCRIPTION Carl Zeiss Stemi DV4 DR 2 8 B 40 003 e 08 00...

Page 19: ...e plate of the stereomicroscope on a flat stable support Remove small cardboard box containing the power supply unit and unpack it Insert primary adapter corresponding to the local line voltage into t...

Page 20: ...h engraved numbers indicating the set overall magnification basic configuration without changeable optical components 5 Focusing knob to enable focusing of the object 6 Illumination control see detail...

Page 21: ...he illumination mode is selected via key M on the illumination control panel This key has been designed as a series switch with 4 switching statuses Each switching procedure between these statuses is...

Page 22: ...p it pressed Transmitted light press keys M and key several times or keep it pressed 3 1 3 3 Specimen protection in transmitted light applications In the Stemi DV4 DR thermally sensitive specimens are...

Page 23: ...apart so that only one light circle is visible on viewing through the eyepieces To search the object set the magnification changing knob 3 4 2 to the lowest magnification first and use the focusing k...

Page 24: ...of the toothed rack If the upper end position is reached the drive will no longer lock in position It is not possible to turn the Stemi carrier out of the stand a cracking noise can be heard if turni...

Page 25: ...crewdriver and remove microscope body from the Stemi carrier in upward direction The ball headed screwdriver is contained in the stand base Pull defective lamp out of the lamp base in axial direction...

Page 26: ...ear edge North and remove it from the stand base Pull defective lamp out of the lamp base in forward direction Make sure not to touch new lamps with your bare hands Cut open the plastic cover of the n...

Page 27: ...ition and use ball headed screwdriver SW 3 to loosen its socket head screw by at least two rotations The ball headed screwdriver is contained in the stand base Hold right focusing knob in position and...

Page 28: ...been loosely inserted Switch on the transmitted light and set maximum intensity If maximum illumination intensity is set in transmitted light the intensity is automatically reduced after 3 minutes to...

Page 29: ...quires the components to be meticulously clean Should the reticle be dirty it must be carefully cleaned first Unscrew both eyepieces 3 13 1 from the eyepiece tubes 3 13 4 Insert reticle 3 13 3 and fie...

Page 30: ...amera Unscrew eyepiece 3 15 5 from one of the eyepiece tubes Attach suitable T2 adapter 3 15 2 to the SLR camera 3 15 1 Screw eyepiece adapter 2 5x T2 3 15 3 into the T2 adapter Insert complete unit c...

Page 31: ...lower edge of the sliding mount should measure 12 mm i e the sliding mount should be attached to the lens mount only until the marking ring Connect supply and signal cables if required If the camera...

Page 32: ...objective filter thread of the DSC F505 camera 3 17 1 until stop Attach sliding mount 3 17 2 with camera 3 17 1 to the lens mount 3 17 4 orient it and use angled screwdriver to tighten the set screw 3...

Page 33: ...led screwdriver to loosen the set screw 3 18 4 on the eyepiece adapter Pull out the lens mount 3 18 5 Unscrew eyepiece 3 18 6 from one of the eyepiece tubes Screw lens mount 3 18 5 optical surface poi...

Page 34: ...t with a lamp voltage close to U max blue LED of the illumination control is blinking remember that the lamp voltage is reduced to constant voltage after three minutes For the operation of the camera...

Page 35: ...piece tubes Insert microhead camera JAI CV M2200 PAL 3 20 1 into the eyepiece adapter for the microhead camera and use angled screwdriver to slightly tighten the clamping screw 3 20 2 Insert eyepiece...

Page 36: ...the image on the monitor is upright and unreversed Tighten retainer ring of the eyepiece adapter and check the setting if required repeat the procedure 3 2 3 7 Eyepiece insertion camera Unscrew eyepie...

Page 37: ...oving in circles starting in the middle optics cleaning solution consisting of 15 isopropanol and 85 medical alcohol benzoline clean glass surface using a slightly moistened wad of cotton by moving in...

Page 38: ...specially authorized personnel To ensure the optimum setting and trouble free function of your microscope even for a longer period of time we would recommend you to conclude a service maintenance con...