6. Nosepiece(1):

  

  The advanced and precise construction of the nosepiece guarantees smooth 

rotation, clear and positive location, and meets the requirement of par-focal and 
par-centred objectives.
7. Mechanical stage(6):

170

160

150

140

130

120

10

0

110

100

50

40

30

20

10

0

0

10

  The stage of this microscope is double 

layers mechanical stage showed in Fig 3. 

Its size is 125(L)X130(W)mm. Push the rod 

of the clamp in the arrow direction showed 
in Fig 3, and insert the apecimen into the 
clamp carefully. Leave your finger away 

from the clamp, the clip will turn back 
slowly, and then the specimen will be 

holded and moved with the clamp. Rotate 

the knob(14)  to move the clamp lengthwise 

in 30mm. Rotate the knob(15)  to move the 
clamp crosswise in 60mm. The moving 
precision is 0.1mm in both directions.

8. ABBE condenser(26)

   The numerical aperture(N.A.) is 1.25. The screw(9) is used to fix the condenser easily 

on the microscope without any tools. The center of the condenser in microscope can be 
adjusted by rotating the black-head screws(8) without any tools as following steps: a)turn 
the objective 4X or 10X in working; b)turn the plate(10) to make the diaphragm diameter 
smaller; c)lower the condenser to make the image of the diaphgram sharp by rotating the 

knob(25); d)rotate the screws(8) to concentre the image of the diaphragm with the 

eyepiece viewing field. The condenser has been adjusted coaxial with the objective 

before the microscope is finished. If not, there will be large different in the veiwing field: 

one side may be dark and another side may be bright. Rotate the knob(25), built-in rack 
and pinion mechanical system controls the condenser up or down. Usually, raise the 
condenser higher when 100X or 40X objective is used; Down the condenser lower when 

10X or 4X objective is used. The highest position of the condenser raising up is limitted to 

lower 0.2mm below the stage surface before the microscope is finished. When it is 
necessary to readjust, do as following steps: a)loose the nut on the screw(12); b)put a 

slide on the stage; c)raise the condenser up to 0.2mm about lower below the slide; 
d)rotate the screw(12) to touch the bottom of the stage; e)fix the screw(12) with the nut. 

 

Fig 3. Mechanical Stage

9. Focusing system:

   It is a coaxial coarse and fine focusing system with rack and pinion mechanism. Its 
focusing range is 20mm, and its precision of fine focusing is 0.01mm. Rotate the coarse 
focusing knob(24) to raise the stage up or down quickly. Rotate the fine focusing knob(23) 
to raise the stage up or down slowly. 

  

Turn the diaphragm plate(10) to adjust the diameter of the iris diaphragm from ? 2 to 

? 30mm to match with the numerical aperture of objective in using. When the diameter of 
the iris diaphragm is 70-80% of the objective's numerical aperture, the image observed is 

sharp in contrast. Look into the tube without eyepiece, you can see the image of the iris 
diaphgram.

  

The filter holder(11) can be turned out to insert the filter when you need. The color of 

the filter may be blue, or green, or yellow.

10. Illumination system:

    Usually, the illuminator of this series microscope is 6V/20W halogen lamp.(It also may 

be 20W incandescent lamp, or 5W fluorescent, or 3.5V/1W LED lamp). Turn the plate
(18) to adjust the brightness. So the image backdrop will be not too bright under the 
lower-powered objectives, or too weak under the higher-powered objectives. LED lamp

Summary of Contents for A11.1303 Series

Page 1: ...www optoecu com 1 15 sale cnoec com A11 1303 Student Biological Microscope Instruction Manual...

Page 2: ...oubleshoot them by yourself according the manual please contact us or our representative in you area 3 Safety Before change a bulb or need to open the base ensure that the microscope has been disconne...

Page 3: ...Focusing System P 5 10 Illumination System P 5 D How to Use and Assemble P 6 E Trouble Shooting P 7 1 Troubles in Operating P 7 2 Troubles in Optical System P 8 3 Troubles in Electric System P 8 F Mai...

Page 4: ...F20X or Huygenian eyepieces as 5X 6X 10X 12 5X 15X 16X The specifications of WF10X are showed as following P 1 a b e c d f Fig 1 The Principle Draft In this microscope the standard outfit of the objec...

Page 5: ...Down 17 Swtich 18 Plate for Adjusting Brightness 16 Screw for Bulb Changing 8 Screw for Centring the Condenser Fig 2 The Structure Draft 3 Mechanical tube length 160mm 4 Conjugated distance between o...

Page 6: ...process software help you to preview measure compare count save and delete Connecting the CCD camera with the TV the micro image will be showed for teaching V type dual observing head 360 rotatable On...

Page 7: ...screen User friendly sophisticated micro image process software help you to preview measure compare count save and delete 3 Insert the CCD camera with the CCD Linker Connecting the CCD camera with th...

Page 8: ...the knob 25 built in rack and pinion mechanical system controls the condenser up or down Usually raise the condenser higher when 100X or 40X objective is used Down the condenser lower when 10X or 4X...

Page 9: ...b D How to use and assemble 1 Unpack the microscope and its parts carefully Check and sort out all parts according to the packing list 2 For the convenience of packaging and delivering the components...

Page 10: ...e 4X and 10X objectives can never come into contacting with your micro slide specimen because of our built in stop The 40X and 100X objectives may occasionally touch the micro slide specimen But becau...

Page 11: ...ld is dark Image moves while focusing The specimen is not caught stably by the clamp The image is yellow Blue filter is not used The viewing field is too dark P 8 Remove the dust and dirt There are du...

Page 12: ...to choose model XSP21 01V or XSP21 01T microscope 1 See Fig 4 and connect the CCD camera with the microscope as following steps 1 Take the plastic dust cover off from the viewing head and screw the v...

Page 13: ...nd its PL pilot lamp is lighting 3 Whether the TV is turned on the mode AV or VIDEO 4 Whether the rod on the right of trinocular head is pushed out 5 Whether the microscope is focused correctly At thi...

Page 14: ...can observe by using common eyepiece at the same time of showing the micro image in the PC screen The electron eyepiece is also used in model XSP21 01M or XSP21 01B microscope See Fig 6 and connect th...

Page 15: ...een as the following ways 1 Turn the plate 19 to make the illuminator darker 2 Turn the plate 10 to make the diameter of the iris diaphragm smaller 3 Rotate the knob 23 to lower the ABBE condenser Whe...

Reviews: