7.6.3.2.2
Spherical Correction – CT Correction
16.
From the Spherical Correction screen tap [CT CORR] to perform a CT correction.
Figure 7-21: Spherical Process
–
CT Correction Screen
17.
Provide a Measured CT value in one of the following ways:
d.
Measure the CT offline and type the value in the “Measured CT” field.
e.
Measure the CT with the onboard part probe by tapping [PROBE CT] to initiate the
automated probing sequence.
Always use caution when using the automated probing sequences to avoid collisions.
18.
Tap [APPLY] to accept the changes and return to the main Spherical Process Corrections screen.
19.
Tap [OK] to accept the new or modified process and return to the HOME screen. Save the
changes to the job by tapping [SAVE JOB].