10
Function
Parameter Settings/Options
Default
Hold Function
C11
0 = turn off hold function
1 = Peak hold - Grabs the highest weight
2 = Manual hold - Grabs the current weight
3 = Auto hold - Automatically holds data when stable
0
Unit Conversion
C12
0 = kg/lb convert on
1 = kg/lb convert off
0
Upper Limit Alarm
C13
Set upper limit within the max. capacity
000000
Lower Limit Alarm
C14
Set lower limit within the max. capacity
000000
Inner Code Display
C15
check the inner code (raw data)
Set Date
C16
Set date from left to right: year/month/day
Set Time
C17
Set the time from left to right: hour/minute/second
Communication
Setting
C18
Set the serial interface data output method:
0 = Turn off serial interface data output
1 = Continuous sending mode, connect remote display
2 = Print mode, connect printer
3 = Command request mode, connect computer.
4 = PC continuous sending mode, connect computer
0
Baud Rate
C19
0=1200 (for remote display)
1=2400
2=4800
3=9600
3
Manual
Zero Range
C20
0 = turn off manually zero setting
1 = ±1% max capacity
2 = ±2% max capacity
4 = ±4% max capacity
10 = ±10% max capacity
20 = ±20% max capacity
100 = ±100% max capacity
10
Initial Zero Range
C21
0 = no initial zero setting
1 = ±1% max capacity
2 = ±2% max capacity
5 = ±5% max capacity
10 = ±10% max capacity
20 = ±20% max capacity
100 = ±100% max capacity
10
Zero Tracking
C22
0= turn off zero tracking
0.5 = ±0.5d d = division
1.0 = ±1.0d
2.0 = ±2.0d
3.0 = ±3.0d
4.0 = ±4.0d
5.0 = ±5.0d
Note: the zero tracking range can not be bigger than
manual zero range
.05
Zero Tracking Time
C23
0 = turn off zero tracking time
1 = 1 second
2 = 2 seconds
3 = 3 seconds
1