background image

9

INDICATOR PARAMETER SETTINGS

The parameter settings menu has a calibration section (C01 to C07) and a  
parameter settings section (C08 and up). 

To enter calibration/parameter settings, follow the procedure below:

1.  Press and hold the POWER 

 

button (1) and the ENTER 

 button (6) at the same time 

2.  Navigate through the settings (C01 to C45) as shown in the table 4 below by using the arrow 

buttons (2-5) and the ENTER button (6) 

3.  Press the 6 

 button to enter/edit the parameter setting

Press the 1 

 button

 to save and exit settings at any time 

Table 1. Indicator Parameter Settings

Function

Parameter Settings/Options

Default

Weighing Unit

C01

1 = kg

2 = lb

1

Decimal Setting

C02

0 = no decimal

1 = 0.0

2 = 0.00

3 = 0.000

4 = 0.0000

0

Graduation Setting

(readability of the least 

significant digit)

C03

options: 1/2/4/10/20/50           

Example with no decimal places (ie. C02=0)

1 = 1 lb

2 = 2 lb

5 = 5 lb

10 = 10 lb

20 = 20 lb

50 = 50 lb

1

Maximum Capacity

C04

set max capacity ex. 100kg = 0100.00

1000

Zero Calibration

C05

0 = zero calibration not needed

1 = set the zero calibration (Please be sure scale is empty 

and the stable light is on) 

0

Calibration

C06

0 = calibration not needed

1 = Ready to calibrate with a calibration weight

0

Restore Default

Settings

C07

0 = do not restore

1 = restore to default settings

0

Warning Tone

C08

0 = turn off warning tone

1 = turn on warning tone

1

Automatic 

Power Off

C09

0 = turn off auto power off

10 = power off automatically if no change within 10 min.

30 = power off automatically if no change within 30 min.

60 = power off automatically if no change within 60 min.

0

Power Saving Mode

C10

LED Version OP900A:

0 = turn off power saving setting

3 = turn off display if no change within 3 minutes

5 = turn off display if no change within 5 minutes

0

Summary of Contents for OP-900-LD

Page 1: ...optimascale com OP 900 LD INDICATOR USER S MANUAL...

Page 2: ...Set Up 1 Features 2 Power Supply 2 Specifications 3 Remote Control 4 Display and Key Descriptions 4 Operating Instructions 5 6 Calibration 7 8 Indicator Parameter Settings 9 11 Helpful Definitions 12...

Page 3: ...e sure the weight is not over the Max capacity as it could damage the load cell inside Material that has a static electric charge could influence the weighing Discharge the static electricity of the s...

Page 4: ...ppm K Sensitivity internal 0 3 V d Input voltage 30 to 30mV DC Excitation circuit 5 VDC 4 wire connection 6 load cell of 350ohm max AC power AC 100 250V use only the included 9V adapter supplied Oper...

Page 5: ...3 SPECIFICATIONS FIGURE 1 INDICATOR MEASUREMENTS 8 5 in 4 75in 2 5in OP 900 LD NET STABLE ZERO lb kg 20 5in 23in 15 25in 3 25 in 6 in...

Page 6: ...ter Set DISPLAY AND KEY DESCRIPTION Net Shows you are in Net weight mode weight without tared weight Stable Shows the weight on the scale is stable Zero Shows you have zeroed the scale lb The weight i...

Page 7: ...l need to tare out the weight Unit Selection To switch between measuring units kg lb press the UNIT 3 button Tare Function The Tare function is used when you only wish to see the current change in wei...

Page 8: ...cumulated number n00 the number of weights you are adding together will flash on the display followed by the total The total will display by flashing between 2 sets of numbers There are 8 digits in to...

Page 9: ...sions on the scale A value of 1 selected and C2 set to 1 the scale will read in 0 1 lb increments 12 Press ENTER to set the value The display will now show C04 13 Press ENTER to access the C4 channel...

Page 10: ...5 The scale will count down from 10 to 0 Once 0 has been reached the display will show CALEnd 26 Press ENTER to continue The display will now show C07 27 Press ON OFF quicly to save and exit the setup...

Page 11: ...ificant digit C03 options 1 2 4 10 20 50 Example with no decimal places ie C02 0 1 1 lb 2 2 lb 5 5 lb 10 10 lb 20 20 lb 50 50 lb 1 Maximum Capacity C04 set max capacity ex 100kg 0100 00 1000 Zero Cali...

Page 12: ...n off serial interface data output 1 Continuous sending mode connect remote display 2 Print mode connect printer 3 Command request mode connect computer 4 PC continuous sending mode connect computer 0...

Page 13: ...ital Filter for filtering moving weight such as animals C28 0 turn off dynamic filter 1 1 digital filter strength 2 2 digital filter strength 3 3 digital filter strength 4 4 digital filter strength 5...

Page 14: ...the time allowed for the scale to fall within the zero tracking range tolerance and still qualify to be zero d Overload Range Weight allowance that is out of the set calibrated range Adds a tolerance...

Page 15: ...w all the anti static rules to avoid damage to your indicator Excitation voltage 5V DC Largest output current 120 mA Excitation circuit 5 VDC 4 wire connection 6 load cell of 350ohm maximum Open the b...

Page 16: ...See Q A section ERR1 During calibration weight is not used or the weight is above the max capacity Use correct weight within the defined range ERR2 During calibration the weight is below the minimum r...

Page 17: ...ing the up arrow key 7 Press the ENTER button to enter your selection 8 The screen will read C21 now 9 Press ENTER key to enter C21 parameter 10 Change the value on the right of C21 to 100 if availabl...

Page 18: ...ascale com for any support related questions Don t forget to visit our website at optimascale com Below are remote guides to print out REMOTE CONTROL Button Sign Function 1 ON OFF 2 Total 3 Unit 4 Tar...

Reviews: