Optima Scale OP-900-LD User Manual Download Page 17

15

Q&A 

Q: The scale does not turn on
A:

Make sure the power cord is plugged in, and that there is power. One easy way to 
test this is by connecting another appliance to the same outlet and see if it’s  
operational

Q: The reading goes negative when a load is applied
A:

Try interchanging the Sig+ and Sig- wiring connected to the load cell and/or  
junction box (if one is used)

Q: How do I resolve ERR6 error?
A:

Please follow the procedure below:
1) Turn on the indicator and make sure nothing is on the scale, and that the scale 
is level and not wobbling
2) Press and hold the “POWER and ENTER” buttons (1+6) simultaneously for 2 
seconds
3) The screen will read “C01” 
4) Using the arrow buttons, change C01 to C20. You have to change the 1st digit 
from 0 to 2 first before you can change the 2nd digit 1 to a 0.
5) Press the “ENTER” button to enter C20 parameter
6) Change the value of C20 on the right to 100 if using the up arrow key
7) Press the “ENTER” button to enter your selection
8)  The screen will read “C21” now
9) Press “ENTER” key to enter C21 parameter
10) Change the value on the right of C21 to 100 if available, 20 if not
11) Press “ENTER” key to enter your selection
12) he screen will read “C22” now
13) Press “POWER” key to save and exit
14) Power the indicator off and then on, and see if this resolves the ERR 6 issue. 
If not, then following the Q&A answers below for resolving “nnnnnn”and “uuuuuu” 
errors

Q:  How do I resolve “nnnnnn” and “uuuuuu” error?
A:

1) Check to see if the cable that runs from the indicator to the junction box is 
damaged. If it is, replace the cable.
2) Open up the junction box (if available) and check to see if there is any water 
damage. If so, replace the junction box
3) Make sure all the wires on all 5 terminal blocks (5 wires on each terminal block) 
are not loose. Re-tighten the screws even if the wires seem to be connected
4) Recalibrate
5) If steps 1-4 do not work, there is a possibility one or more load cells are  
defective (consult with [email protected] for further instructions)

Summary of Contents for OP-900-LD

Page 1: ...optimascale com OP 900 LD INDICATOR USER S MANUAL...

Page 2: ...Set Up 1 Features 2 Power Supply 2 Specifications 3 Remote Control 4 Display and Key Descriptions 4 Operating Instructions 5 6 Calibration 7 8 Indicator Parameter Settings 9 11 Helpful Definitions 12...

Page 3: ...e sure the weight is not over the Max capacity as it could damage the load cell inside Material that has a static electric charge could influence the weighing Discharge the static electricity of the s...

Page 4: ...ppm K Sensitivity internal 0 3 V d Input voltage 30 to 30mV DC Excitation circuit 5 VDC 4 wire connection 6 load cell of 350ohm max AC power AC 100 250V use only the included 9V adapter supplied Oper...

Page 5: ...3 SPECIFICATIONS FIGURE 1 INDICATOR MEASUREMENTS 8 5 in 4 75in 2 5in OP 900 LD NET STABLE ZERO lb kg 20 5in 23in 15 25in 3 25 in 6 in...

Page 6: ...ter Set DISPLAY AND KEY DESCRIPTION Net Shows you are in Net weight mode weight without tared weight Stable Shows the weight on the scale is stable Zero Shows you have zeroed the scale lb The weight i...

Page 7: ...l need to tare out the weight Unit Selection To switch between measuring units kg lb press the UNIT 3 button Tare Function The Tare function is used when you only wish to see the current change in wei...

Page 8: ...cumulated number n00 the number of weights you are adding together will flash on the display followed by the total The total will display by flashing between 2 sets of numbers There are 8 digits in to...

Page 9: ...sions on the scale A value of 1 selected and C2 set to 1 the scale will read in 0 1 lb increments 12 Press ENTER to set the value The display will now show C04 13 Press ENTER to access the C4 channel...

Page 10: ...5 The scale will count down from 10 to 0 Once 0 has been reached the display will show CALEnd 26 Press ENTER to continue The display will now show C07 27 Press ON OFF quicly to save and exit the setup...

Page 11: ...ificant digit C03 options 1 2 4 10 20 50 Example with no decimal places ie C02 0 1 1 lb 2 2 lb 5 5 lb 10 10 lb 20 20 lb 50 50 lb 1 Maximum Capacity C04 set max capacity ex 100kg 0100 00 1000 Zero Cali...

Page 12: ...n off serial interface data output 1 Continuous sending mode connect remote display 2 Print mode connect printer 3 Command request mode connect computer 4 PC continuous sending mode connect computer 0...

Page 13: ...ital Filter for filtering moving weight such as animals C28 0 turn off dynamic filter 1 1 digital filter strength 2 2 digital filter strength 3 3 digital filter strength 4 4 digital filter strength 5...

Page 14: ...the time allowed for the scale to fall within the zero tracking range tolerance and still qualify to be zero d Overload Range Weight allowance that is out of the set calibrated range Adds a tolerance...

Page 15: ...w all the anti static rules to avoid damage to your indicator Excitation voltage 5V DC Largest output current 120 mA Excitation circuit 5 VDC 4 wire connection 6 load cell of 350ohm maximum Open the b...

Page 16: ...See Q A section ERR1 During calibration weight is not used or the weight is above the max capacity Use correct weight within the defined range ERR2 During calibration the weight is below the minimum r...

Page 17: ...ing the up arrow key 7 Press the ENTER button to enter your selection 8 The screen will read C21 now 9 Press ENTER key to enter C21 parameter 10 Change the value on the right of C21 to 100 if availabl...

Page 18: ...ascale com for any support related questions Don t forget to visit our website at optimascale com Below are remote guides to print out REMOTE CONTROL Button Sign Function 1 ON OFF 2 Total 3 Unit 4 Tar...

Reviews: