IFU_EN_GEN-HT Version 1.01
July2020
10
Technical Specification
Sample Number
12 x 8 wells
Sample Volume
10 µl to 150 µl
Touchscreen
10.1” High-brightness LCD with multi-touch projective capacitive touch
panel (1280 x 800)
Heater technology
Ceramic substrate with resistive coating
Cooling method
Forced convection
Temperature sensors
High-precision thermistor
Temperature control type
Multi-zone independent digital PID
Temperature control range
ambient - 100°C
Temperature accuracy
±0.1°C
Temperature uniformity
across block
±0.2°C
Temperature gradient
Programmable up to 8°C
Optics source
470 nm & 590nm dual colour LED with high-quality interference filter 40 nm
band pass
Detection optics
Photodiodes with high-quality interference filters 510-560 nm band pass &
620 nm long pass
Operating temperature
0°C - 40°C
Storage Temperature
20°C - 70°C
Approvals
CE
Dimensions
635mm (L) X 434mm (W) X 153mm (H)
Weight
18kg / 40lbs
Connections
1 x USB ‘B’
2 x USB ‘A’
1 x Power in
1 x RJ45 – Network connection
1 x USB ‘A’ – diagnostic port
Environmental protection
IP20
Wireless Connections
Bluetooth & WiFi
Power supply
Input: 100-240V AC, 50/60Hz
RoHS Compliance
OptiGene Limited declares that, to the best of our knowledge, all electrical and electronic equipment (EEE) sold by the company
are in compliance with Directive 2011/65/EU of the European Parliament and of the Council of 8 June 2011 on the restriction of the
use of certain hazardous substances in electrical and electronic equipment (also known as “RoHS2”).
REACh Statement
OptiGene Limited also declares that no products sold by the company are known to contain any of the substances of very high
concern (SVHC) as intentionally added components with regards to the European Regulation No. 1907/2006 (Registration,
Evaluation, Authorisation & restriction of Chemicals).