INST No.INE-459-0P2CE Ver. 1.11
- 18 -
8.1 Modes at start up
The following operation modes are available by the key combinations at the start up.
Keys Modes
Remarks
Press
MEAS
key only
(about 1 second)
Standard measurement or
continuous measurement
Measurement with the same measurement
mode as the last starting up
Press
MEAS
while
pressing
△
key
Standard measurement
Press
MEAS
key
while pressing
▽
key
Continuous measurement
The " " mark will light under the main
marker "
CONT
".
8.2 Table of screens
The screens displayed in the external display are the following 3 kinds basically.
Screen Outline
Measurement
screen
Measurement by
MEAS
key
In the hold mode in the standard or continuous measurement mode, the setting and
automatic calculation of emissivity rate is possible.
Data storage
screen
Data storage number, stored data display and storing measured data into memory are
possible. By pressing
MEM
key in the standard or continuous measurement mode,
the mode will become the data storage mode and the " " mark will light under
the main marker "
MEM
".For returning to the standard or continuous measurement
mode, press
MEM
key again.
Measurement
mode settings
screen
Display and setting of measurement mode. By pressing
SEL
key for about 2
seconds in the standard measurement mode or continuous measurement mode, the
mode will become the measurement mode settings mode.
In the measurement mode settings mode, by pressing
SEL
key for about 2 seconds
or any key is not pressed for 1 minute, the mode will return to the measurement
mode or the data storage mode.
8.3 Measurement mode settings/display
8.3.1 Emissivity rate setting/display
Parameter item
Display
Parameter
Initial
Paragraph
Emissivity rate (ratio)
ε
(
ε
r)
1.900 to 0.100
1.000
5.4
*
The emissivity rate can not be set in the data storage mode.
8.3.2 Data storage number setting/display
Parameter item
Display
Parameter
Initial
Paragraph
Data storage number
NO
1 to 100
1
6.2
*
The data storage number is only displayed in the data storage mode.
8.4 System settings
Settings Display Parameter
Initial Paragraph
Low temperature alarm setting
AL
oFF, 400 to 3000ºC/752 to 5432ºF oFF 5.6.1
High temperature alarm setting
AH
oFF, 400 to 3000ºC/752 to 5432ºF oFF 5.6.2
Signal modulation mode selection modu dELy,
PEAk
dELy
5.5.1
Modulation value setting *
tAu
dEc
0.0, 0.2, 0.5, 1.0 (second)
0, 2, 5, 10ºC (ºF)/second
0.0s
0ºC/s
5.6.3
All stored data erasing
AdEL
no, yES
no
6.4
Measuring unit selection
Unit C(ºC),
F(ºF)
C(ºC)
5.5.2
2-color/single color wide
selection
CoLr 2,
1
2 5.5.3
* The mode for the modulation value setting differs by the selection of the signal modulation mode.
(Ref: [5.5.1 Signal modulation mode selection])
8. List of Starting Up Modes