- 6 -
(Common to [5.1 Standard measurement mode] and [5.2 Continuous measurement mode])
External display
Main marker
Internal display
Status marker
Main display
Sub marker Sub display
•
Data in the sub display (Common to [5.1 Standard measurement mode] and [5.2 Continuous
measurement mode])
Sub marker Data
displayed in sub display
ε
(
ε
r) Emissivity
rate
(
ε
) for single color type, Emissivity ratio (
ε
r
) for 2-color type
MIN
Minimum temperature during the
MEAS
key is being pressed
MAX
Maximum temperature during the
MEAS
key is being pressed
AVE
Average temperature during the
MEAS
key is being pressed (Moving average of
25 points)
NO
Stored data storage number. (Lights in the data storage mode only.)
5.1 Standard measurement mode
This mode is for measurement with holding the thermometer by hands.
• Sight through the viewfinder and match the center circle of the targeting circle to the center of
object measured.
• Press
MEAS
key for about 1 second to turn on the power supply and start a measurement.
The temperature measured will be displayed in the internal and external displays, and the status
marker “
MEAS
” will light in the external display.
(The measurement is continued during
MEAS
key is being pressed.)
• Select a data to be displayed in the sub display by pressing
SEL
key when you need.
By releasing
MEAS
key, the measurement will stop and the measured value will be held.
The status marker "
MEAS
" will go off and "
HOLD
" will light.
5.2 Continuous measurement mode
5.2.1 Start of continuous measurement
This mode is for continuous measurement by fixing the thermometer on a tripod or a universal head.
• Sight through the viewfinder and match the center circle of the targeting mark to the center of
object measured.
• For the continuous measurement mode, press
MEAS
key while pressing
▽
key.
When all segments of the external display (Ref. 3.3 [External display]) light, release
MEAS
key
quickly before a beep sounds. The " " mark under the main marker "
CONT
" and the status
marker "
HOLD
" will light.
・
5. Measuring
*
Make sure to press any keys until you hear an audible beep.
Tb CONT MEM PEAK
MEAS
HOLD AL
AH
C
F
MAX
AVE
ε
r
MIN
TC
NO
• "
oFL
" will be displayed in the main and sub displays if the temperature data
measured is higher than the measuring range or "
uFL
" will be displayed if it is
lower than the measuring range.
[Temperature displaying "
oFL
"] 2-color: 2020°C (3668°F) Single color: 3020°C (5468°F)
[Temperature displaying "
uFL
"] 2-color: 580°C (1076°F) Single color: 380°C (716°F)
R e m a r k s
• The internal display will go off in 10 seconds in the hold mode and the power
supply will be automatically turned off if any key is not pressed for 15 seconds in
the hold mode.
R e m a r k s
• The status marker "
MEAS
" will light depending on its releasing timing of
MEAS
key and the continuous measurement may start.
In this case, press
MEAS
key only again to light the status marker "
HOLD
" so that
the continuous measurement is temporally stopped.