INST No.INE-459-0P2CE
Ver. 1.11
- 10 -
5.5.3 2-color type/single color type selection
Select a measurement mode by 2-color type or single color type in the VF-3000.
• Press
MEAS
key for about 1 second to turn on the power supply and then release
MEAS
key. The status
marker "
HOLD
" will light. Then press
SEL
key for about 2 seconds in the hold mode to move to the setting
mode. A setting value will be displayed in the main display and its mode will be displayed in the sub display.
• Press
SEL
key to display "
CoLr
" in the sub display.
• By pressing
△
or
▽
key, either "
2
" (2-color type) or
"
1
" (single color type) will blink in the main display.
• Select either one and press
ENT
key to store it. The “
2
”
(2-color type) or “
1
” (single color type) selected will stop
blinking and light in the main display.
5.6 Measurement mode settings
(This paragraph explains the items highlighted.)
Measurement modes are shown in the list below. By referring to the list, set the desired modes.
Settings/Mode selections
Sub display
Setting range
Initial Paragraph
Low temperature alarm setting
AL
oFF, 400 to 3000ºC/752 to 5432ºF
oFF
5.6.1
High temperature alarm setting
AH
oFF, 400 to 3000ºC/752 to 5432ºF
oFF
5.6.2
Signal modulation mode selection modu dELy,
PEAk
dELy
5.5.1
Modulation value setting
*
tAu
dEc
0.0, 0.2, 0.5, 1.0 (second)
0, 2, 5, 10ºC (ºF)/second
0.0s
0ºC/s
5.6.3
All stored data erasing
AdEL
no, yES
no
6.5
Measuring unit selection
Unit
C (ºC), F (ºF)
C
5.5.2
2-color/single color selection
CoLr
2, 1
2
5.5.3
* The mode for the modulation value setting differs by the selection of the signal modulation mode.
(Ref: [5.5.1 Signal modulation mode selection])
5.6.1 Low temperature alarm setting
This setting is for a judgment of low alarm during measurement. When the low alarm is judged, the status
marker "
AL
" will light and the buzzer will sound. When the "
oFF
" is selected, neither the alarm judgment can
be executed nor the buzzer can be sound.
• Press
MEAS
key for about 1 second to turn on the power supply and then release
MEAS
key. The status
marker "
HOLD
" will light. Then press
SEL
key for about 2 seconds in the hold mode to move to the setting
mode. A setting value will be displayed in the main display and its mode will be displayed in the sub display.
• Press
SEL
key to display "
AL
" in the sub display. By pressing
△
key, the least significant digit will blink
to set it.
• Set the desired figure by
△
or
▽
key and press
ENT
key for shifting the setting digit to the next higher digit.
• Repeat the above procedure up to the most significant
digit.
• By pressing
ENT
key at the most significant digit, the
setting value will be stored.
5. Measuring
*
Make sure to press any keys until you hear an audible beep.
Tb CONT MEM PEAK
MEAS
HOLD
AL
AH
C
F
MAX
AVE
ε
r
MIN
TC
NO
Main display
Sub display
• The initial setting is "
2
" (2-color type).
R e m a r k s
• If
SEL
key is pressed for 2 seconds in the setting mode or if any key is not pressed for
1 minute, the thermometer will return to the measurement mode.
R e m a r k s
• If
SEL
key is pressed for 2 seconds in the setting mode or if any key is not pressed for
1 minute, the thermometer will return to the measurement mode.
R e m a r k s
[How to reset the low temperature alarm to "oFF"]
• Press
SEL
key for about 2 seconds to display "
AL
" on the sub display. By changing the
displayed numeric figure to be lower than 399ºC (751ºF) by
△
,
▽
and
ENT
keys,
"
oFF
" will be displayed.
• Press
ENT
key to store it.
Reference
• The setting range is “oFF” or 400 to 3000ºC/752 to 5432ºF. The initial is "
oFF
".
Reference
Main display
Sub display
Tb CONT MEM PEAK
MEAS
HOLD
AL
AH
C
F
MAX
AVE
ε
r
MIN
NO