background image

20

21

 When both right and left arrow buttons             are pressed simultaneously, Signal 

Output of 20mA will be produced irrespectively of the measurement value. 

CAUTION:

Fine Adjustment of Upper Limit 20mA

2

2

1

While making sure of the values displayed by the 
connecting equipment such as recorder, make the Fine 
Adjustments of the Upper Limit 20mA with the up and 
down arrow buttons           .

Simultaneously press both right and left 
arrow buttons            .
At this time, "-20-" is indicated in the 
Display. And, "ZERO" comes on and 
"SPAN" blinks.

4

Press         to terminate the Fine Adjustments of the Upper 
Limit 20mA.

Reference 

To cancel the Fine Adjustment of the Upper Limit 20mA, press         .

At this time, after making sure that the indicated values and 
measured values by the official method have linear relationship 
as shown below, calculate a Span-Adjusted  value using 
displayed values and measured values as close as possible to 
the Span-Adjusted value.

Examine the correlation between values measured by the official 
method and the values indicated by SC-T4 at a place where the 
instrument is installed.

Indicated values

Indicated values

1

2

The following calculating formula is used.
Span setting value = [Present Span-Adjusted value] x Measured 
value (official method)/Indicated value (SC-T4).
Notes: The Span-Adjusted value set 1000 as a factory setting.

3

The Span-Adjusted value is changed into the following value by 
referring to page 12([5] Operation (3) 0(ZERO)/Span-Adjustments).
When the indicated value is 800 mg/l and the measured value 
according to official method is 600mg/L:
Span-Adjusted value =1000x600/800=750

In the case of the Present Span-Adjusted value is 1000.

4

8

Measured v

alues b

y  

the official method

Measured v

alues  b

the official method

When Err4 or Err5 is displayed, the value is out of the setting range. The 

available setting range of Span-Adjusted value is from 200 to 3000 mg/l.

CAUTION:

7

Correction

Depending on the measuring environment, indicated value may differ 
from the value by the official method. In this occation, correct the 
indicated value using the Span-Adjustment function according to the 
following procedure. 

Summary of Contents for SS Checker SC-T4

Page 1: ...is instruction manual should be kept for future reference such as maintenance Transmitter Instruction Manual 2007 7 59 1362 1 For more information on OPTEX products contact your dealer or visit our we...

Page 2: ...ture reference such as maintenance Transmitter Instruction Manual 0 1 Table of Contents 1 For safe use 2 External Features 3 Installation Installation Wiring Extension of Detecter Cable 4 Operation Pa...

Page 3: ...ve gas is in danger of intruding through the piping etc into the Transmitter Do not bind the signal output cable with the power cable or do not put them in the one Cable Clamp Install the transmitter...

Page 4: ...t On Transmitter Operation panel Display 2 External Features Cover Cable Clamps Key Hook Lock Lever Power Switch Circuit Breakers Terminal Block Operation Buttons Display Terminal Cover Alarm Level In...

Page 5: ...luding wiring resistance For connection to the Self Checking Relay Output terminal and Alarm Relay Output terminal load resistance should be 240 VAC and 1A Max Although a protective circuit is built i...

Page 6: ...o carry out Fine Adjustment of Signal Output Low Limit 4mA Press the upper and lower arrow buttons simultaneously to carry out Fine Adjustment of Signal Output High Limit 20mA Move the digit to be ent...

Page 7: ...d water or ion exchange water 1 Clean the Detector and the Detector Windows Every time the MODE button is pressed the Display is changed into each setting To change the setting value select the item a...

Page 8: ...s not 1000 use to change the Display into 1000 indication When the display shows 1000 press to terminate the 0 Span Adjustments 8 Press for 2 seconds or more SPAN blinks on the Display Pip Pip Pip 4 A...

Page 9: ...ayed the value is out of the setting range Alarm Timer can be set for 1 to 120 minutes or OFF Setting unit is 1 minute The Alarm Timer is set for OFF as a factory setting Alarm Timer Setting Ex To set...

Page 10: ...d lower arrow buttons button increases a numeric value and button decreases the value While pressing buttons the values displayed are continuously changed To cancel the Signal Output Low Limit Setting...

Page 11: ...Pip Pip Fine Adjustment of Signal Output 6 Fine Adjustment of Lower Limit 4mA and Upper Limit 20mA of Signal Output 4 20mA can be made When both right and left arrow buttons are pressed simultaneously...

Page 12: ...es measured by the official method and the values indicated by SC T4 at a place where the instrument is installed Indicated values Indicated values 1 2 The following calculating formula is used Span s...

Page 13: ...indicated on the Display Err 5 is indicated on the Display The Detector is out of order A Detector Cable is not wired properly The Detector is out of order A Detector Cable is not wired properly The D...

Page 14: ...Power Supply for Sensor 12 VDC Signal Output analog 4 20mA resistance load of 300 or less Self checking Relay Output non voltage C contact capacity 240 VAC 1A resistance load Alarm Relay Output non vo...

Reviews: