background image

24

25

Maintenance

Periodic inspection

Calibration

Long-term storage

When the Cover or the Display is soiled, first wipe away lightly with a clean soft 
cloth damped by diluted mild detergent solution and then wipe off moisture with a 
dry clean soft cloth. 

Inspect the following items every 3 months:
   

 Make sure that the Transmitter is fixed securely.

   

 Make sure that the Transmitter is not damaged.

   

 Make sure that the screws of the Terminal Block are not rusted.

The SC-T4 is designed so as to make measurements stably for a long time. In 
order to maintain the reliability of measurement, however, carry out calibration at 
least once a year (Refer to page 10 ([5] Operation (1) Calibration)).

When the SC-T4 is not used for a prolonged period, turn off the Power Switch 
and disconnect the Power Cable from the power supply prior to safekeeping.

10

Maintenance

CAUTION:

 Do not wipe the transmitter with solvent such as Thinner.

Calibration Output

Power Supply for Sensor (12 VDC)

Signal Output (analog 4-20mA, resistance load of 300

Ω

 or less)

Self-checking Relay Output (non-voltage C-contact  capacity 240 VAC, 1A resistance load)

Alarm Relay Output (non-voltage C-contact  capacity 240 VAC, 1A resistance load)

Detector signal input (analog 4-20mA, input resistance approx.100

Ω

)

Self-checking input

IEC 60529  /  IP65 (jetproof type)

AC100 

 240V

±

10%    50/60Hz

Power supply

Dimensions

SC-T4

0.1mg/l (0-49.9mg/l), 1mg/l (50-1000mg/l)

Output

Power 

consumption

Normal: 8VA or less, during washing: 16VA or less

 (including detector, at the time of analog signal input/output of 20mA)

Display resolution

Input (detector)

The specifications herein are subject to change without prior notice due to improvements.

11

Specifications

12

Dimensions

[mm]

16

38

75

34

34

34

34

220

Installation pitch 186

Installation pitch 125

240

(Max. 190)

162

Installation hole 

φ

9

Applicable cable 
diameter

φ

6~8

Hole for 
padlock

φ

7

Model No.

Main material

Operating temperature

-20 to +50 

 , humidity 95%Rh or less (Avoid direct sunlight)

Polycarbonate

Degree of protection

HxWxD = approx.162x240x75 mm

Weight

Approx. 1.6kg (including the Mounting Bracket)

Summary of Contents for SS Checker SC-T4

Page 1: ...is instruction manual should be kept for future reference such as maintenance Transmitter Instruction Manual 2007 7 59 1362 1 For more information on OPTEX products contact your dealer or visit our we...

Page 2: ...ture reference such as maintenance Transmitter Instruction Manual 0 1 Table of Contents 1 For safe use 2 External Features 3 Installation Installation Wiring Extension of Detecter Cable 4 Operation Pa...

Page 3: ...ve gas is in danger of intruding through the piping etc into the Transmitter Do not bind the signal output cable with the power cable or do not put them in the one Cable Clamp Install the transmitter...

Page 4: ...t On Transmitter Operation panel Display 2 External Features Cover Cable Clamps Key Hook Lock Lever Power Switch Circuit Breakers Terminal Block Operation Buttons Display Terminal Cover Alarm Level In...

Page 5: ...luding wiring resistance For connection to the Self Checking Relay Output terminal and Alarm Relay Output terminal load resistance should be 240 VAC and 1A Max Although a protective circuit is built i...

Page 6: ...o carry out Fine Adjustment of Signal Output Low Limit 4mA Press the upper and lower arrow buttons simultaneously to carry out Fine Adjustment of Signal Output High Limit 20mA Move the digit to be ent...

Page 7: ...d water or ion exchange water 1 Clean the Detector and the Detector Windows Every time the MODE button is pressed the Display is changed into each setting To change the setting value select the item a...

Page 8: ...s not 1000 use to change the Display into 1000 indication When the display shows 1000 press to terminate the 0 Span Adjustments 8 Press for 2 seconds or more SPAN blinks on the Display Pip Pip Pip 4 A...

Page 9: ...ayed the value is out of the setting range Alarm Timer can be set for 1 to 120 minutes or OFF Setting unit is 1 minute The Alarm Timer is set for OFF as a factory setting Alarm Timer Setting Ex To set...

Page 10: ...d lower arrow buttons button increases a numeric value and button decreases the value While pressing buttons the values displayed are continuously changed To cancel the Signal Output Low Limit Setting...

Page 11: ...Pip Pip Fine Adjustment of Signal Output 6 Fine Adjustment of Lower Limit 4mA and Upper Limit 20mA of Signal Output 4 20mA can be made When both right and left arrow buttons are pressed simultaneously...

Page 12: ...es measured by the official method and the values indicated by SC T4 at a place where the instrument is installed Indicated values Indicated values 1 2 The following calculating formula is used Span s...

Page 13: ...indicated on the Display Err 5 is indicated on the Display The Detector is out of order A Detector Cable is not wired properly The Detector is out of order A Detector Cable is not wired properly The D...

Page 14: ...Power Supply for Sensor 12 VDC Signal Output analog 4 20mA resistance load of 300 or less Self checking Relay Output non voltage C contact capacity 240 VAC 1A resistance load Alarm Relay Output non vo...