background image

14

15

Alarm Level can be set for 1 to Span-Adjusted value or OFF.

Setting unit is 1 mg/l.

The Alarm Level is set for OFF as a factory setting.

Alarm Level Setting

Ex: To set the Alarm Level for 30 mg/l ;

4

Press         to terminate the Alarm Level Setting.

3

Enter "30" on the Display.

2

1

Press         for 2 seconds or more, the 
Alarm Level Indicator blinks. 

Press         and select the Alarm Level. 

Pip

Pip Pip

Move the digit to be entered using the right and left arrow buttons             
(1 digit       10 digit       100 digit       1000 digit). 
At this time, the selected digit blinks.

 

 

 

Reference 

Enter a numerical value (0, 1, 2, ..., 9) into the selected digit using the 
upper and lower arrow buttons            .        button increases a numeric 
value and          button decreases the value.
While pressing           buttons, the values displayed are continuously 
changed.

Reference:

To cut off the Alarm Relay Output, enter "0". Then "OFF"appears on the Display. Next, 

press         .

To cancel the Alarm Level Setting, press         .

CAUTION:

 When Err6 is displayed, the value is out of the setting range.

Alarm Timer can be set for 1 to 120 minutes or OFF.

Setting unit is 1 minute.

The Alarm Timer is set for OFF as a factory setting.

Alarm Timer  Setting

Ex: To set the Alarm Timer for 60 minutes ;

3

Enter "60" on the Display.

2

1

Press         for 2 seconds or more, the 
Alarm Timer Indicator blinks. 

Press         and select the Alarm Timer. 

Pip

Pip Pip

 

To change the Display for the Measurements Indication mode, press        
again.

4

Press         to terminate the Alarm Timer Setting.

Move the digit to be entered using the right and left arrow buttons              
(1 digit       10 digit       100 digit). 
At this time, the selected digit blinks.

 

 

 

Reference 

Enter a numerical value (0, 1, 2, ..., 9) into the selected digit using the 
upper and lower arrow buttons            .        button increases a numeric 
value and          button decreases the value.
While pressing           buttons, the values displayed are continuously 
changed.

Reference:

To cut off the Alarm Timer, enter "0". Then "OFF"appears on the Display. Next, 

press         .

To cancel the Alarm Timer Setting, press         .

CAUTION:

 When Err6 is displayed, the value is out of the setting range.

 

To change the Display for the Measurements Indication mode, press        
again.

Summary of Contents for SS Checker SC-T4

Page 1: ...is instruction manual should be kept for future reference such as maintenance Transmitter Instruction Manual 2007 7 59 1362 1 For more information on OPTEX products contact your dealer or visit our we...

Page 2: ...ture reference such as maintenance Transmitter Instruction Manual 0 1 Table of Contents 1 For safe use 2 External Features 3 Installation Installation Wiring Extension of Detecter Cable 4 Operation Pa...

Page 3: ...ve gas is in danger of intruding through the piping etc into the Transmitter Do not bind the signal output cable with the power cable or do not put them in the one Cable Clamp Install the transmitter...

Page 4: ...t On Transmitter Operation panel Display 2 External Features Cover Cable Clamps Key Hook Lock Lever Power Switch Circuit Breakers Terminal Block Operation Buttons Display Terminal Cover Alarm Level In...

Page 5: ...luding wiring resistance For connection to the Self Checking Relay Output terminal and Alarm Relay Output terminal load resistance should be 240 VAC and 1A Max Although a protective circuit is built i...

Page 6: ...o carry out Fine Adjustment of Signal Output Low Limit 4mA Press the upper and lower arrow buttons simultaneously to carry out Fine Adjustment of Signal Output High Limit 20mA Move the digit to be ent...

Page 7: ...d water or ion exchange water 1 Clean the Detector and the Detector Windows Every time the MODE button is pressed the Display is changed into each setting To change the setting value select the item a...

Page 8: ...s not 1000 use to change the Display into 1000 indication When the display shows 1000 press to terminate the 0 Span Adjustments 8 Press for 2 seconds or more SPAN blinks on the Display Pip Pip Pip 4 A...

Page 9: ...ayed the value is out of the setting range Alarm Timer can be set for 1 to 120 minutes or OFF Setting unit is 1 minute The Alarm Timer is set for OFF as a factory setting Alarm Timer Setting Ex To set...

Page 10: ...d lower arrow buttons button increases a numeric value and button decreases the value While pressing buttons the values displayed are continuously changed To cancel the Signal Output Low Limit Setting...

Page 11: ...Pip Pip Fine Adjustment of Signal Output 6 Fine Adjustment of Lower Limit 4mA and Upper Limit 20mA of Signal Output 4 20mA can be made When both right and left arrow buttons are pressed simultaneously...

Page 12: ...es measured by the official method and the values indicated by SC T4 at a place where the instrument is installed Indicated values Indicated values 1 2 The following calculating formula is used Span s...

Page 13: ...indicated on the Display Err 5 is indicated on the Display The Detector is out of order A Detector Cable is not wired properly The Detector is out of order A Detector Cable is not wired properly The D...

Page 14: ...Power Supply for Sensor 12 VDC Signal Output analog 4 20mA resistance load of 300 or less Self checking Relay Output non voltage C contact capacity 240 VAC 1A resistance load Alarm Relay Output non vo...

Reviews: