 Before making the 0-Adjustment, be sure to clean the Detector Windows.

 Before making the 0-Adjustment, be sure to carry out calibration.

 After making the 0-Adjustment, be sure to make the Span-Adjustment.

0 (ZERO) / Span - Adjustments

CAUTION:

Be sure to carry out calibration before setting these operations.

4

3

After cleaning the Detector Windows, immerse the Detector 
in distilled water or ion-exchange water.

5

7

Cover one of the Detector Windows completely to 
be light-shielded for 30 seconds or more. 

6

2

Press         and select the 0-Adjustment. 

1

After accustoming the Detector to water 
temperature for 5 minutes or so, make sure that 
air bubbles are not attached on the Detector 
Windows. And press         for 2 seconds or more, 
"ZERO" blinks on the Display.

When the Display is not "0", use              to change the Display into 
"0" indication. When the display shows "0", press 

.

Pip

Press         and select the Span-Adjustment. 

Pip Pip

CAUTION:

 When Err4, Err5, or Err6 is displayed, the SC-T4 is out of the setting range, or the 

Detector is not immersed in clean water during the 0-Adjustment, or the Detector 
Windows may not completely be light-shielded during the Span-Adjustment.

When the Display is not "1000", use            to change the 
Display into "1000" indication. When the display shows "1000", 
press         to terminate the 0 / Span -Adjustments.

8

Press         for 2 seconds or more, "SPAN" blinks 
on the Display.

Pip

Pip Pip

4

Alarm Level, Alarm Timer, Signal Output Response Time, 
Signal Output Range

The Alarm Level Indicator appears when the measurements 
get higher than the Alarm Level. Both the Alarm Timer 
indicator and Alarm Level Indicator blink when the 
measurements get longer and higher than the setting value 
of the Alarm Timer and Alarm Level. And in this case, Alarm 
Relay Output will be produced. 

When the measurements get higher than the Alarm Level continuously 
beyond the time set by the Alarm Timer, the Alarm Relay Output will be 
produced.

Although the measurements get 
higher than the Alarm Level, since 
higher period is shorter than that of 
the Alarm Timer, Alarm Relay 
Output will not be produced.

When the measurements 
become lower than the 
Alarm Level, the Alarm 
Relay Output is released.

(Higher)

Alarm Relay Output

Alarm Level

(Lower)

Alarm Timer

Alarm Timer

time

The Alarm Relay Output will be produced. 
While the measurements values are 
higher than the Alarm Level, the Alarm 
Relay Output will continue to be produced.

Alarm Relay Output Setting

1

These setting is set for OFF as a factory setting.
Make setting as occasion demands.

Ex: At the time of 
      Alarm Relay Output

Summary of Contents for SS Checker SC-T4

Page 1: ...is instruction manual should be kept for future reference such as maintenance Transmitter Instruction Manual 2007 7 59 1362 1 For more information on OPTEX products contact your dealer or visit our we...

Page 2: ...ture reference such as maintenance Transmitter Instruction Manual 0 1 Table of Contents 1 For safe use 2 External Features 3 Installation Installation Wiring Extension of Detecter Cable 4 Operation Pa...

Page 3: ...ve gas is in danger of intruding through the piping etc into the Transmitter Do not bind the signal output cable with the power cable or do not put them in the one Cable Clamp Install the transmitter...

Page 4: ...t On Transmitter Operation panel Display 2 External Features Cover Cable Clamps Key Hook Lock Lever Power Switch Circuit Breakers Terminal Block Operation Buttons Display Terminal Cover Alarm Level In...

Page 5: ...luding wiring resistance For connection to the Self Checking Relay Output terminal and Alarm Relay Output terminal load resistance should be 240 VAC and 1A Max Although a protective circuit is built i...

Page 6: ...o carry out Fine Adjustment of Signal Output Low Limit 4mA Press the upper and lower arrow buttons simultaneously to carry out Fine Adjustment of Signal Output High Limit 20mA Move the digit to be ent...

Page 7: ...d water or ion exchange water 1 Clean the Detector and the Detector Windows Every time the MODE button is pressed the Display is changed into each setting To change the setting value select the item a...

Page 8: ...s not 1000 use to change the Display into 1000 indication When the display shows 1000 press to terminate the 0 Span Adjustments 8 Press for 2 seconds or more SPAN blinks on the Display Pip Pip Pip 4 A...

Page 9: ...ayed the value is out of the setting range Alarm Timer can be set for 1 to 120 minutes or OFF Setting unit is 1 minute The Alarm Timer is set for OFF as a factory setting Alarm Timer Setting Ex To set...

Page 10: ...d lower arrow buttons button increases a numeric value and button decreases the value While pressing buttons the values displayed are continuously changed To cancel the Signal Output Low Limit Setting...

Page 11: ...Pip Pip Fine Adjustment of Signal Output 6 Fine Adjustment of Lower Limit 4mA and Upper Limit 20mA of Signal Output 4 20mA can be made When both right and left arrow buttons are pressed simultaneously...

Page 12: ...es measured by the official method and the values indicated by SC T4 at a place where the instrument is installed Indicated values Indicated values 1 2 The following calculating formula is used Span s...

Page 13: ...indicated on the Display Err 5 is indicated on the Display The Detector is out of order A Detector Cable is not wired properly The Detector is out of order A Detector Cable is not wired properly The D...

Page 14: ...Power Supply for Sensor 12 VDC Signal Output analog 4 20mA resistance load of 300 or less Self checking Relay Output non voltage C contact capacity 240 VAC 1A resistance load Alarm Relay Output non vo...

Reviews: