![background image](http://html.mh-extra.com/html/omron/zx2-series/zx2-series_user-manual_1633407059.webp)
Thickness
59
ZX2 User’s Manual
INTRODUCTION
BASIC
SETUP
PREPARATION
FOR
MEASUREMENT
TROUBLE-
SHOOTING
INDEX
CONTENTS
SETTING
TRANSITION
CHARTS
FLOW OF
OPERATION
MAIN
APPLICATIONS
& SETTING
METHODS
DETAILED
SETTINGS
SPECIFI-
CATIONS
Height
Double
Sheet
Detection
Positioning
Eccentricity
and Surface
Deflection
Steps
and
Warpage
Thickness
8
Mutual interference prevention
9
Threshold Setting
Button
Operation
Display
Description of Operation
Explanation of
Selection Menu
Press the button on the CH1
Amplifier Unit to display
.
Press the button to display
.
Press the
button to apply
the mutual interference
prevention setting.
Button
Operation
Display
Description of Operation
Explanation of
Selection Menu
Press the button on the CH2
Amplifier Unit to display the
HIGH threshold.
(Setting example)
Non-defective product
thickness 3 to 8 mm
Set the MAX and MIN
thicknesses to be regarded
as OK to the HIGH and LOW
thresholds, respectively.
* If the button is pressed
when the cursor is at the
right-most digit or the
button is pressed when the
cursor is at the left-most digit,
the setting will be canceled.
* Set so that the HIGH
threshold is greater than the
LOW threshold.
Press the button to enable
setting of the HIGH threshold.
Set any value.
Press the
button to move
the digit, press the button to
change the numeric value, and
set the HIGH threshold.
Press the
button to apply
the setting.
Press the button to display
the LOW threshold.
Press the button to enable
setting of the LOW threshold.
Set any value.
Press the
button to move
the digit, press the button to
change the numeric value, and
set the LOW threshold.
Required
Set this item on the CH1 Amplifier Unit to
prevent the influence of mutual
interference between two Sensor Heads.
Press to display
.
Default value: OFF
Press to display.
Required
Set the range of measured values to be judged
as PASS on the CH2 Amplifier Unit by setting
the HIGH and LOW thresholds.
Press to display
.
Lit
L
H
P
3
8
NG
OK
NG
Press to set.
[Change numeric value]
[Move digit]
Press to display
.
Lit
Press to set.
[Change numeric value]
[Move digit]