No.14S056-11
22
STC-CMC120APCL, STC-CMB120APCL
Specifications and Users guide
4.4 AOI Output Timing
This camera can be set the AOI up to 8.
Variable Partial Start Line(VAS*), Variable Partial Effective Line(VAH*), Horizontal Start
Position(HAS*), Horizontal Effective Pixel(HAW*) can be set for each eight regions. One region can
be set on horizontal direction. And vertical area can not be overlapped for each region.
As follows setting is the example of four area of interest (AOI)
AOI_
A
AOI_
B
AOI_
C
AOI_
D
VAH
A
A0-A1h
HAS
A
B0-B1h
VAS
A
90-91h
VAH
B
A2-A3h
VAS
B
92-93h
VAH
C
A4-A5h
VAS
C
94-95h
VAH
D
A6-A7h
VAS
D
96-97h
HAS
B
B2-B3h
HAS
C
B4-B5h
HAS
D
B6-B7h
4096
3072
HAW
A
C0-C1h
HAW
B
C2-C3h
HAW
C
C4-C5h
HAW
D
C6-C7h
For more information please contact Aegis Electronic Group, Inc. *(888)687-6877 *[email protected] *http://www.aegiselect.com
Aegis
Electronic
Group,
Inc.