However, this is a case that has been used in the recommended environment, such as regular
maintenance and inspection, there may be a difference through the use situation arises.

Matters relating to the maintenance and inspection

[User, and skilled in the art by the maintenance and inspection matters]

Notes

- For a while when you re-use the equipment that was not used, make sure that you always the
equipment before use to operate successfully and safely.

- Body, accessories, connector, in the opening portion of the housing, with blowing liquid Ketari,
pouring, please do not spill. Otherwise, you may get an electric shock.

- Disposable products in order to prevent infection, do not re-use. (There is also a thing that can
be re-used only the same person to be measured.)

- Do not use paint thinner, solvents such as benzene to clean. These solvents will damage the
surface of the plastic.

Inspection before use

- In appearance, there is no deformation such as by dropping, Ri was dirty, Please check not wet.

Cuff, please make sure that you are using the specified ones.

Inspection after the power is turned on

- Smoke or out, or there is no strange smell, please check whether the abnormal noise is not
heard.

[Maintenance and inspection matters by the trader]

Notes

- Clean, because there is a risk of electric shock, please be performed by far the communication
connector. 

* Please insert the communication connector After cleaning from the well dried. 

• The unit, autoclave, can not be used gas sterilization (formaldehyde gas, EOG, high
concentration ozone, etc.). The equipment will cause a deterioration.

Inspection before use

· That the cords are not damaged (exposure of the core, disconnection, etc.), 
Please make sure that there is no slack in the cords of connection.

- Please check do not place heavy objects on top of the code.

Inspection after the power is turned on

• Press the each switch, make sure that the function is activated.

• Each measurement is performed normally, please make sure the measurement error is within
the reference value.

Summary of Contents for M3500

Page 1: ...ressure module Blood pressure measurement module M3500 Be sure to refer to the instruction manual Creation or revision date March 1 2012 third edition October 17 2011 second edition Approval and notification etc Brand Name Blood pressure measurement module M3500 Attached document management code 22200BZX00227000_A_01 Approval Number 22200BZX00227000 Approval date March 2010 General designation for...

Page 2: ...e Magnetic resonance imaging apparatus There is a risk of causing accidents caused by combination induced electromotive force of the MRI devices how to use That you do not use If the exterior of the cuff there is corruption There is a risk of rupture during the measurement Shape and structure and principles etc This unit is a blood pressure measuring module designed to measure the non invasive blo...

Page 3: ...ured the pulsation of the artery as the vibration of the cuff pressure to measure the blood pressure It pressed enough pressure to the cuff wound to the upper arm and the blood flow will stop but the beat of the artery is present also vibration cuff internal pressure in response to it Then when gradually reduce the cuff pressure vibration of the cuff pressure is gradually increases eventually it r...

Page 4: ...tion check Method of operation Wrap the cuff on the arm of the subject Power is supplied to the unit and then to the blood pressure measurement possible state Measurement start and press the stop switch to start the measurement After the measurement remove the cuff from the subject PRECAUTIONS In the case of the following or please carefully apply to the following of the subject The direction from...

Page 5: ... transfusion please do not wound the cuff to a limb that has been vein secured Please do not wound the cuff to the limb with an artificial dialysis shunt Please measure with an appropriate cuff size Doing so may result of measurement error From time to time to check the measurement site please change the measurement site if there is an abnormality Or causing the peripheral circulatory failure rash...

Page 6: ...priate it can cause burns in the subject mounting portion Please carefully read the instruction manual of the electric knife for more information Microwave therapy equipment equipment that generates the noise such as you might not be able to if measured correctly in the vicinity Such as the commercially available personal computer to the external I O connector please do not direct connection with ...

Page 7: ...ch as benzene to clean These solvents will damage the surface of the plastic Inspection before use In appearance there is no deformation such as by dropping Ri was dirty Please check not wet Cuff please make sure that you are using the specified ones Inspection after the power is turned on Smoke or out or there is no strange smell please check whether the abnormal noise is not heard Maintenance an...

Page 8: ...t wipe the electrical connector portion Packing Packed in a single unit Manufacturing distributors and manufacturers of name and address etc Name or name Omron Healthcare Co Ltd First class medical equipment manufacturing and sales Address etc Yubinbango617 0002 Kyoto Muko Terato cho Kyunotsubo 53 address phone number Customer Support Center 0120 103 203 24 hours a day toll free Foreign manufactur...

Reviews: