Blood pressure measurement module M3500
Page 4 of 8
https://translate.googleusercontent.com/translate_c?depth=1&hl=…03%3Fview%3Dbody&usg=ALkJrhhqzooj9yOkvRDfve8nZaQansOoVg#WARNING
2.
3.
Four.
Five.
(1)
(2)
(3)
(Four)
1.
Error average of the auscultation: ± 5mmHg within the standard deviation: 8mmHg within
Cuff pressure display of error: within ± 3mmHg
Quick exhaust
Adult (children and infants):
The time required in order to rapidly exhaust to 15mmHg from 260mmHg: 10 seconds or less
newborn:
The time required in order to rapidly exhaust to 5mmHg from 150mmHg: less than 5 seconds
Cuff pressure display of stability:
10,000 change in the display value of the cycle cuff pressure after the simulated measurement of:
3mmHg below
** Maximum pressure
Adults: 300mmHg
Newborn: 150mmHg
Method of operation or use method and the like
Since that is operated by the operation buttons of the assembled multi-item monitoring, operation
button of the device is not used.
Method of operation is as follows for the operation check.
Method of operation
Wrap the cuff on the arm of the subject.
Power is supplied to the unit, and then to the blood pressure measurement possible state.
Measurement start and press the stop switch, to start the measurement.
After the measurement, remove the cuff from the subject.
PRECAUTIONS
In the case of the following, or please carefully apply to the following of the subject.
- The direction from the doctor has been pointed out the bleeding tendency or hypercoagulable
state (can cause faulty circulation by point-like bleeding or blood clots after tightening the arm in
the cuff.)
And natural rubber (latex) the subject of allergy: the cuff being attached genus or option on this
device may have the use of natural rubber. Please check whether there is an abnormality in the
site where the hose touches. Natural rubber, itching, redness, nettle rash, swelling, fever,
breathing difficulties, asthma-like symptoms, decrease in blood pressure, may cause rare allergic
symptoms such as shock or be. In such a case that caused the symptoms, please immediately
subjected to a mid-sealed appropriate measures to use. The presence or absence of natural rubber
use mosquito off the document attached Please check