1.

Error average of the auscultation: ± 5mmHg within the standard deviation: 8mmHg within

Cuff pressure display of error: within ± 3mmHg

Quick exhaust

Adult (children and infants): 
The time required in order to rapidly exhaust to 15mmHg from 260mmHg: 10 seconds or less
newborn: 
The time required in order to rapidly exhaust to 5mmHg from 150mmHg: less than 5 seconds

Cuff pressure display of stability:

10,000 change in the display value of the cycle cuff pressure after the simulated measurement of:
3mmHg below

** Maximum pressure

Adults: 300mmHg 
Newborn: 150mmHg

Method of operation or use method and the like

Since that is operated by the operation buttons of the assembled multi-item monitoring, operation
button of the device is not used. 
Method of operation is as follows for the operation check.

Method of operation

Wrap the cuff on the arm of the subject.

Power is supplied to the unit, and then to the blood pressure measurement possible state.

Measurement start and press the stop switch, to start the measurement.

After the measurement, remove the cuff from the subject.

PRECAUTIONS

In the case of the following, or please carefully apply to the following of the subject.

- The direction from the doctor has been pointed out the bleeding tendency or hypercoagulable
state (can cause faulty circulation by point-like bleeding or blood clots after tightening the arm in
the cuff.)

And natural rubber (latex) the subject of allergy: the cuff being attached genus or option on this
device may have the use of natural rubber. Please check whether there is an abnormality in the
site where the hose touches. Natural rubber, itching, redness, nettle rash, swelling, fever,
breathing difficulties, asthma-like symptoms, decrease in blood pressure, may cause rare allergic
symptoms such as shock or be. In such a case that caused the symptoms, please immediately
subjected to a mid-sealed appropriate measures to use. The presence or absence of natural rubber
use mosquito off the document attached Please check

Summary of Contents for M3500

Page 1: ...ressure module Blood pressure measurement module M3500 Be sure to refer to the instruction manual Creation or revision date March 1 2012 third edition October 17 2011 second edition Approval and notification etc Brand Name Blood pressure measurement module M3500 Attached document management code 22200BZX00227000_A_01 Approval Number 22200BZX00227000 Approval date March 2010 General designation for...

Page 2: ...e Magnetic resonance imaging apparatus There is a risk of causing accidents caused by combination induced electromotive force of the MRI devices how to use That you do not use If the exterior of the cuff there is corruption There is a risk of rupture during the measurement Shape and structure and principles etc This unit is a blood pressure measuring module designed to measure the non invasive blo...

Page 3: ...ured the pulsation of the artery as the vibration of the cuff pressure to measure the blood pressure It pressed enough pressure to the cuff wound to the upper arm and the blood flow will stop but the beat of the artery is present also vibration cuff internal pressure in response to it Then when gradually reduce the cuff pressure vibration of the cuff pressure is gradually increases eventually it r...

Page 4: ...tion check Method of operation Wrap the cuff on the arm of the subject Power is supplied to the unit and then to the blood pressure measurement possible state Measurement start and press the stop switch to start the measurement After the measurement remove the cuff from the subject PRECAUTIONS In the case of the following or please carefully apply to the following of the subject The direction from...

Page 5: ... transfusion please do not wound the cuff to a limb that has been vein secured Please do not wound the cuff to the limb with an artificial dialysis shunt Please measure with an appropriate cuff size Doing so may result of measurement error From time to time to check the measurement site please change the measurement site if there is an abnormality Or causing the peripheral circulatory failure rash...

Page 6: ...priate it can cause burns in the subject mounting portion Please carefully read the instruction manual of the electric knife for more information Microwave therapy equipment equipment that generates the noise such as you might not be able to if measured correctly in the vicinity Such as the commercially available personal computer to the external I O connector please do not direct connection with ...

Page 7: ...ch as benzene to clean These solvents will damage the surface of the plastic Inspection before use In appearance there is no deformation such as by dropping Ri was dirty Please check not wet Cuff please make sure that you are using the specified ones Inspection after the power is turned on Smoke or out or there is no strange smell please check whether the abnormal noise is not heard Maintenance an...

Page 8: ...t wipe the electrical connector portion Packing Packed in a single unit Manufacturing distributors and manufacturers of name and address etc Name or name Omron Healthcare Co Ltd First class medical equipment manufacturing and sales Address etc Yubinbango617 0002 Kyoto Muko Terato cho Kyunotsubo 53 address phone number Customer Support Center 0120 103 203 24 hours a day toll free Foreign manufactur...