Blood pressure measurement module M3500
Page 3 of 8
https://translate.googleusercontent.com/translate_c?depth=1&hl=…03%3Fview%3Dbody&usg=ALkJrhhqzooj9yOkvRDfve8nZaQansOoVg#WARNING
(2)
Four.
Five.
6.
7.
1.
Classification of equipment
** Blitz degree of protection against by the mounting portion of the classification: arm
band B-shaped mounting portion
usage environment
Temperature: 0 ~ 50
℃
Humidity: 30 to 85% (non-condensing)
Standard accessory
The following as a standard accessory comes with one point.
** · IEC adult cuff or GS cuff M
** - Adult Kafuhosu (3.5m)
Silicon tube chassis joint / nut
Optional Accessories
Apart from the standard accessories, to this device can be combined with the product has the
following.
· IEC adult oversized cuff · IEC children cuff (9cm, 7cm)
** - Adult cuff (S, M, L, XL)
** · GS cuff (SS, S, L, XL)
And neonatal / infant Disupokafu (No.10, No.11, No.12, No.13)
** - Adult Kafuhosu (1.5m)
And neonatal for Kafuhosu (3.5m)
Air filter
Operating and operation principle
In the oscillometric method blood pressure measurement oscillometric method, it captured the
pulsation of the artery as the vibration of the cuff pressure, to measure the blood pressure. It
pressed enough pressure to the cuff wound to the upper arm and the blood flow will stop, but the
beat of the artery is present, also vibration cuff internal pressure in response to it. Then, when
gradually reduce the cuff pressure vibration of the cuff pressure is gradually increases,
eventually it reached a peak,
In addition go and vibration by reducing the cuff pressure would be Ikima smaller. The
relationship between the increase and decrease of the vibration of the cuff pressure and cuff
pressure in this series of processes and memory and calculation, to determine the blood pressure
value.
Systolic blood pressure cuff internal pressure of typically, when vibration is rapidly increased,
And the lowest blood pressure cuff internal pressure at the time of the decrease rapidly.
In addition, the cuff pressure when the vibration of the peak is the mean arterial pressure.
Intended use, efficacy or effect
This device, by the non-invasive measurement of arterial blood pressure is used to display the
systolic and diastolic blood pressure.
Item specifications, etc.
Error of blood pressure measurement by the clinical performance test