Blood pressure measurement module M3500
Page 5 of 8
https://translate.googleusercontent.com/translate_c?depth=1&hl=…03%3Fview%3Dbody&usg=ALkJrhhqzooj9yOkvRDfve8nZaQansOoVg#WARNING
Important Precautions
In the periphery of, the apparatus and the person to be measured, the power supply, such as
mobile phones and small radio (PHS terminal that is allowed to use is excluded) because it is not
concerned with the cut again.
· Unmeasurable and when you have doubts about the measurements, please make sure the person
to be measured first.
· Please use devices also document attached to the cuff to be used from read them carefully to
measure. In this package document it does not has indicated up to those notes.
Specifying the consumables, please do not use other than the optional accessories.
- For each person to be measured, or please make sure that the alarm settings are appropriate
before use.
[Non-invasive blood pressure measurement]
Such as, infusion and transfusion, please do not wound the cuff to a limb that has been vein
secured.
- Please do not wound the cuff to the limb with an artificial dialysis shunt.
· Please measure with an appropriate cuff size. Doing so may result of measurement error.
- From time to time, to check the measurement site, please change the measurement site if there
is an abnormality. Or causing the peripheral circulatory failure, rash caused by sweat, and cause
internal bleeding.
- Please note the bending of the Kafuhosu. Peripheral dysfunction due to the fact that not exhaust
may occur.
• If the following can not be unmeasurable or measured correctly.
- Peripheral circulatory failure and significant hypotension, the case of hypothermia
• If a high frequency of arrhythmia
• If there is a weak continuous vibration from the heart massage in and outside
- Convulsions, tremors, involuntary movements, if there is a body movement such as
trembling
• If the cuff mounting portion is different from the height of the heart (10cm is different
when the blood pressure value height might maximum 7 ~ 8mmHg different)
• If you talk or move during the measurement
• If you attach the cuff from the top of the thick clothing