65
Applied part
Type BF (cuff)
Protection against
electric shock
Internally powered ME equipment
Class II ME equipment (AC adapter)
Maximum temperature
of the applied part
Lower than +118.4 °F (+48 °C)
Battery Life Information
The battery life is based on 2 days (48 hours) with the following use: in
23°C environment, 200 mm wrist circumference, 8 times of blood pressure
measurement (inflates to 150 mmHg) per day, 16-hour step counting per
day, 8-hour sleep measurement per day, 96 times of button pressing per
day, 16 times of incoming call per day, 1 time of ringing alarm per day, 16-
hour Bluetooth communication per day.
The battery life depends on the terms of use and the environment of storage.
Blood pressure measurement
Cuff pressure range
0 to 299 mmHg
Blood pressure
measurement range
SYS: 60 to 230 mmHg
DIA: 40 to 160 mmHg
Pulse measurement
range
40 to 180 beats / min.
Accuracy
Pressure: ±3 mmHg
Pulse: ±5 % of display reading
Inflation
Automatic by electric pump
Deflation
Automatic rapid deflation
Measurement method
Oscillometric method
11. Specifications