30
1) File Management Controls
There are functions for loading images and saving images.
Pushing the “ Load Image
“
button allows to show the saved images.
Select the image files from “ Blood Flow Image
“
or “ Real Image
“
folders.
The whole display can be saved by “ Save Display Image
“
button. This
function is the same as the “ Print Screen function
“
in Windows. The folder for
saving appears on the display, then enter the file name. The image is saved in
the “ Analyzed Data
“
folder.
Pushing the “ Save Image
“
button allows to save one color image displayed on
the screen.
Select “ New Analysis
“
from the top menu bar for the new analysis. The
“ Analysis Configuration File
“
is shown again.
The example of the file name of a saved image is like
“" Image HR BF R3T
〜
3500No001" .
The letters stand for :
Image : the original name of the images saved by LSI,
HR : measurement mode,
BF : blood flow image,
R3T : the round number (this example is the third measurement),
〜
3500 : time of saved image from the start of the saving,
No001 : the number of the image.
Be careful not to change the file name. Change the whole file names in one
folder, Real Images and BF Images, when the original name is changed. If the
original names are different in one folder, the images are not recognized as the
same sequence image data.
2) Real Image Display
The Real Images by laser light are shown in black and white.
3) Blood Flow Image Display
The Blood Flow Images are shown in color according to the blood flow value.
The color of the blood flow values are assigned by the color bar. Two types of
color bars can be selected. One is “ Standard
“
, and the other is “ Enhanced
“
.
They are in Tool in the menu bar.
4) MAX, MIN
This function is to change the color gradation of the color bar to show the
suitable blood flow images.
MAX : the maximum value of the color bar
MIN: the minimum value of the color bar
8a