23
values of the axes. The maximum number of X-axis is 10,000. The slider bar at
the bottom of the graph can be used to display a specific time (number) in the
graph..
13) Blood Flow Histogram
Histogram of the number of pixels corresponds to blood blow values in ROIs.
The X-axis shows blood flow value in ROI, and Y-axis shows the number of
pixels at a blood flow value on the X-axis.
14) Area (pixels)
his column shows the number of pixels in ROI1 and ROI2.
15) Marker
Pushing this button makes the punch in “ Blood Flow Image “ display, and
makes the line in “ Blood Flow Plot 12) “.
This is used to mark the withe punches during the measurement. The marks can
be seen in LIA display, and saved in the Excel file as “ M “.
16) Saving the setting and initial setting
(1) Pushing “ Save Setting “ button will save the setting of
(a) Values of “ MAX “ and “ MIN “
(b) Measurement mode
(c) AVG number for HS-AVG mode
(d) Frame Interval time
(e) N / Time
(f) Repeat Interval time
(g) N of Times
(h) Y-axis of “ Blood Flow Plot “
(u) Y-axis of “ Blood Flow Histogram “
(j) Figures and the positions of “ ROI “s
(k) Lens Coefficient and Laser Distance
(l) “ AUTO “ or “ MANUAL “ of Gain & Intensity
(2) Pushing “ Initial Setting “ button 16) clears the all saved settings, and the
setting will be the initial condition.
17) Menu bar
(1) File
(a) Capture Screen : The whole display can be save in the clip board.
This saved image can be used to attach on some documents.
(b) New Experiment : For a new measurement. It is recommended to make
the new folder for every new measurements.
(c) Save Images : Same function as “ Save Image “ button in 4).