Series TCL-3M165E2 

Temperature Calibrator  

Summary of Contents for TCL-3M165E2 Series

Page 1: ...Series TCL 3M165E2 Temperature Calibrator...

Page 2: ...Series TCL 3M165E2 2...

Page 3: ...5 4 2 Electrical connection 16 4 3 Preparing of the calibrator 17 4 3 1 Dry block calibrator 18 4 3 2 Calibrator for infrared thermometer 19 4 3 3 Calibrator for surface temperature sensors 20 4 3 4 M...

Page 4: ...tenance and cleaning 50 8 1 Maintenance 50 8 2 Cleaning 51 8 3 Recalibration 52 8 4 Adjustment 53 9 Decommissioning and disposal 54 10 Technical data 55 10 1 Characteristics TCL 3M165E2 55 10 2 Heatin...

Page 5: ...ndicates dangers which could lead to serious health defects or to death WARNING CAUTION Risk of injury This sign indicates dangers that cause personal injuries that can lead to health defects or cause...

Page 6: ...l and laboratory tasks Thus for example thermometers temperature switches thermostats resistance thermometers and thermo elements can be directly connected and checked Versions The TCL 3M165E2 is avai...

Page 7: ...rature calibrators are delivered in special protective packaging Save the packaging for returning the instrument safely to the manufacturer for recalibration or repair Scope of delivery and accessorie...

Page 8: ...injuries Never touch the metal block the tank the adapter sleeve or the test specimen at temperatures above 35 C or below 10 C 95 F or below 50 F Allow the calibrator to cool before you remove the tes...

Page 9: ...This can be based on training or relevant instructions Personnel must be aware of this operating manual and have access to it at all times The electrical connection should only be carried out by a ful...

Page 10: ...he test sample is securely fixed in the calibrator Use only suitable adapter sleeves or calibration inserts When doing so also ensure that the structural safety of the calibrator is retained Expert mo...

Page 11: ...ise excessive limescale and soiling will build up in the calibrator tank Silicone oil Use only the silicone oil recommended here Always read the safety data sheet supplied with the silicone oil before...

Page 12: ...heat insulated calibration block with heating or cooling elements and an integrated sensor for the reference temperature In the case of the dry block and multifunction calibrator this is a metal block...

Page 13: ...the calibrator They give the test specimen a secure fit and ensure optimum heat transfer Once all the preparations have been carried out the calibrator can be switched on 4 4 The calibrator heats or c...

Page 14: ...defined in the factory in which the basic settings of the calibrator are stored If desired customer specific test tasks can also be pre set These protected test tasks can be neither deleted nor edited...

Page 15: ...nt conditions p 56 4 1 Operating conditions Select a safe installation site for commissioning the calibrator Installation site and operating position Only suitable for indoor use do not use outdoors O...

Page 16: ...with the approved supply voltage 10 1 Ensure that the power voltage is the same as that specified on the type plate Please follow the local regulations of the energy supplier Connect the calibrator o...

Page 17: ...e tank must be emptied and cleaned to prevent the adapter sleeve or the calibration insert from getting stuck Empty and clean the tank before you insert another adapter sleeve or calibration insert Me...

Page 18: ...ment element of the test specimen must be located in the homogeneous temperature zone of the adapter sleeve Inserting The appropriate adapter sleeve is inserted into the metal block with the aid of th...

Page 19: ...d insert Inserting The infrared insert is inserted into the metal block with the aid of the sleeve exchange tool External reference sensor optional Align the infrared insert so that the hole for the e...

Page 20: ...rface insert and the metal block This prevents the surface insert from getting stuck in the metal block IMPORTANT NOTICE The calibration of surface temperature sensors is very difficult and not uncont...

Page 21: ...d with a safety valve which is activated once the pressure reaches 1 5 bar 21 75 psi This can result in hot steam being released Always unscrew the transport cover before putting the micro bath into s...

Page 22: ...liquids supply varying calibration results due to their specific characteristics Adjustment to the respective calibration liquid has to be carried out by the manufacturer In order to achieve the best...

Page 23: ...of the calibration liquid causes contamination and can damage the calibrator Ensure that the maximum fill level is not exceeded during operation The fill level in the tank or tub insert rises as a res...

Page 24: ...sleeve exchange tool only for tub insert Place the magnetic stirrer in the tank tub insert Insert the sensor cage Insert the test specimens in the sensor cage This accounts for the volume of the sens...

Page 25: ...port or storage After transport storage or long periods of non use moisture can seep into the heating elements magnesium oxide For drying the calibrator must be slowly heated up During this process th...

Page 26: ...d Switch off the calibrator only when the safe temperature range has been reached Press the key until the main window is displayed Press the key to switch off the calibrator The calibrator regulates t...

Page 27: ...tive window of the parameter is opened or displayed The windows Calibrator Set Up and Select measurement report are opened with the icons and The window Select Test Task is opened from the parameter f...

Page 28: ...he saved test tasks and create or configure your own test tasks p 36 Switching off By using the icon you can switch the calibrator off 4 4 Switching on cool down and switching off Display range The re...

Page 29: ...mind From the windows Select Function 5 4 1 and Select DUT 5 5 1 you can switch to the expert mode for administering and configuring Follow the instructions in the corresponding sections Changes to f...

Page 30: ...displays the elements available in the window This can be available test tasks recorded measurement logs saved test specimens or functions of the calibrator The desired element is selected by tapping...

Page 31: ...ry is generated for the selection list and the relevant window is displayed Delete The selected entry in the selection list is deleted Search The selection list is searched for an entry for example p...

Page 32: ...d function by tapping on in the selection list The information on the function is immediately displayed in the information box From the toolbar you can accept the selected function or switch to expert...

Page 33: ...ap on the parameter field of the desired parameter The parameter s window is opened You can configure the parameter there Name You can change the name of the function here After tapping the virtual ke...

Page 34: ...urned to the previous window 5 5 Test specimen DUT Device under test Here you can select and manage existing test specimens You can create new test specimens or tailor unprotected test specimens to ma...

Page 35: ...window Configure DUT is opened Details pertaining to the window can be found in the following section 5 5 3 Configure DUT From the Manage DUT window you can use the icon to go to the window Configure...

Page 36: ...e icon to search for a test task A virtual keyboard with an input line is displayed Input the desired search term and confirm your input The result of the search is shown in the selection list Manage...

Page 37: ...information box Select the desired type and confirm the input The new type of data acquisition is adopted and you will be returned to the previous window When the data acquisition is activated the te...

Page 38: ...the selection of the Behavior at test end p 40 individual parameter fields have another designation or are suppressed The information box has several parameter fields for setting the test points All t...

Page 39: ...The new values are adopted the interval is calculated afresh and you will be returned to the previous window Cycles You can set the number of cycles here For every cycle the test points in the select...

Page 40: ...the parameter field Safety temperature value Input the respective safety temperature value and confirm the input The new value is adopted and you will be returned to the previous window 5 6 6 Alarm s...

Page 41: ...There the relevant information is displayed or the parameters are configured Parameter fields of the Calibrator Set Up window System information The most important data of your calibrator is displaye...

Page 42: ...subnet mask are requested from the network The two parameter fields are disabled In addition for Manual Input the IP address from the numeric keypad of the window and confirm the input with the icon...

Page 43: ...lash drive Insert the USB flash drive in the calibrator and wait for 20 30 s so that the USB flash drive is recognized Now tap on the parameter field Software update Follow the instructions on the scr...

Page 44: ...flash drive in the calibrator and wait for 20 30 s so that the USB flash drive is recognized Now tap the parameter field Import user files Follow the instructions on the screen Export user files You...

Page 45: ...the desired measurement log by tapping on in the selection list The information on the measurement log is immediately displayed in the information box Delete measurement report On tapping the icon th...

Page 46: ...with 4 1 the electrical connections have been made correctly 4 2 the inside of the block and the surface of the adapter sleeve or the calibration insert are dry You can safely remove existing ice or c...

Page 47: ...of the test task Display range The following information is shown in the display range Current temperature values of the reference used Current test point Probable remaining test duration for the cur...

Page 48: ...to OMEGA ENGINEERING Costs incurred due to inadequate cleaning of the device and possible costs for disposal and or personal injuries will be billed to the operating company WARNING Risk of injury due...

Page 49: ...ial attention to sealing grooves and slits Attach a note describing the malfunction state the application field and the chemical physical properties of the media Please follow the instructions on the...

Page 50: ...ors check the filled height of the calibration liquid When doing so note the Information on filling amounts 4 3 4 2 Annually Subject all the parts of the calibrator to a visual inspection for corrosio...

Page 51: ...tor to OMEGA ENGINEERING for repairs 7 7 1 8 2 Cleaning External cleaning Clean the TCL 3M165E2 with a dry or slightly damp lint free cloth Do not use sharp objects or aggressive agents for cleaning E...

Page 52: ...ank Remove the magnetic stirrer with the help of the magnetic lifter Dry the tank the sensor cage the magnetic stirrer and the drain syringe thoroughly Silicone oil Remove the sensor cage from the tan...

Page 53: ...ment values can be freely selected You can equalize the calibrator directly on site with your standard thermometer and thus achieve a maximum of display accuracy It is not necessary to send it to the...

Page 54: ...mpletely cooled down 4 4 Switching on cool down and Decommissioning Remove all connected sensors and devices Switch off the calibrator and disconnect the power plug Empty the calibrator tank in case t...

Page 55: ...0 5 C 0 9 F 1 C 1 8 F 0 1 C 0 18 F Emission degree Infrared 0 9994 Stability of the control Dry block Infrared Surface Micro bath 0 010 C 0 018 F 0 020 C 0 018 F 0 150 C 0 18 F 0 010 C 0 018 F Measur...

Page 56: ...m 6 6 ft Process variables Metal block Tank opening Borehole Depth 60 mm 2 36 in 170 mm 6 69 in Dimensions Width Height Depth 210 mm 8 27 in 380 50 mm 14 96 1 97 in 300 mm 11 81 in Weight 13 kg 28 66...

Page 57: ...he heating up and cooling times are dependent on the parameters of the test task used and the ambient conditions The following times are guide values for a room temperature of 23 C 73 4 F without sett...

Page 58: ...For your notes 58 For your notes...

Page 59: ...Series TCL 3M165E2 Technical changes reserved 59...

Page 60: ...Series TCL 3M165E2 60 M 5663 0318...