11.- ADJUSTMENT AND CALIBRATION PROCEDURE

11.1.- RTD MODELS

A decade box or RTD simulator is required.

1.- Check that the instrument is not powered.
2.- Connect the RTD simulator or decade box to the signal input connector (See paragraph 8.3).
3.- Remove the rear cover taking off all the screws to obtain access to the control board.

(See page 7 paragraph 7).

4.- Power up the instrument with the appropriate power supply and wait until 3 minutes before to

start the adjust.

5.- Adjust the RTD simulator to generate the low signal indicated in TABLE-A.
6.- Display must shows the DISPLAY value indicated in TABLE-A, if not turn the trimmer PT1 located

on the Signal Conditioner (See paragraph 10.1) until display shows the DISPLAY value.

7.- Adjust the RTD simulator to generate the high signal, see TABLE-B.
8.- Display must shows the DISPLAY value indicated in TABLE-B, if not turn the trimmer PT2 located

on the Signal Conditioner (See paragraph 10.1) until display shows the DISPLAY value.

9.- Repeat steps 5 to 8 until the two values are correct.

10.- Check that the linearity remains correct, testing some other points of the range.

11.2.-  THERMOCOUPLE MODELS

A Thermocouple simulator with Cold Junction compensation is required.

1.- Check that the instrument is not powered.
2.- Connect the Thermocouple simulator to the signal input connector (See paragraph 8.4).
3.- Remove the rear cover taking off all the screws to obtain access to the control board.

(See page 7 paragraph 7).

4.- Power up the instrument with the appropriate power supply and wait until 3 minutes before to

start the adjust.

5.- Adjust the Thermocouple simulator to generate the low signal indicated in TABLE-A.
6.- Display must shows the DISPLAY value indicated in TABLE-A, if not turn the trimmer PT1located

on the Signal Conditioner (See paragraph 10.2) until display shows te correct value.

7.- Adjust the Thermocouple simulator to generate the high signal indicated in TABLE-B.
8.- Display must shows the DISPLAY value, if not turn the trimmer PT2 located on the Signal

Conditioner (See paragraph 10.2) until obtain the DISPLAY value.

9.- Repeat steps 5 to 8 until the low and the high signals are correct.

10.- Check that the linearity remains correct, testing some other points of the range.

  

                          TABLE - A

   

TABLE - B

MODEL

 LOW SIGNAL

DISPLAY

MODEL

 HIGH SIGNAL

DISPLAY

  

 M1

0°C

   000.0

    M1

      600°C

   600.0

   M2

0°F

   0000

    M2

      500°F

   0500

   M3

0°F

   000.0

    M3

      200°F

   200.0

    J1

0°C

   0000

     J1

      600°C

   0600

    J2

        32°F

   0032

     J2

    1000°F

   1000

   K1

0°C

   0000

    K1

    1200°C

   1200

   K2

        32°F

   0032

    K2

    2000°F

   2000

   T1

0°C

   0000

    T1

      400°C

   0400

   T2

0°F

   0000

    T2

      400°F

   0400

   E1

0°C

   0000

    E1

      600°C

   0600

   E2

        32°F

   0032

    E2

    1200°F

   1200

   R1

      300°C

   0300

    R1

    1700°C

   1700

   R2

        32°F

   0032

    R2

    3100°F

   3100

   S1

      200°C

   0200

    S1

    1700°C

   1700

   S2

        32°F

   0032

    S2

    3100°F

   3100

   L1

          0°C

   0000

    L1

      600°C

   0600

Summary of Contents for LDP-144 Series

Page 1: ...LDP 124 AND LDP 144 SERIES FOR TEMPERATURE http www omega com e mail info omega com User s Guide...

Page 2: ...ree 0800 1 66342 e mail czech omega com France 9 rueDenisPappin 78190Trappes Tel 33 130 621 400 Fax 33 130 699 120 Toll Free in France 0800 4 06342 e mail france omega com Germany Austria Daimlerstras...

Page 3: ...COUPLE R 6 LDP 1X4 L1 TEMPERATURE INDICATOR FOR THERMOCOUPLE L 6 GENERAL SPECIFICATIONS 5 7 FRONT VIEW DESCRIPTION 6 7 REAR VIEW DESCRIPTION 7 7 WIRING 8 8 POWER SUPPLY RECOMMENDED WIRING 8 1 8 FUSES...

Page 4: ...and control wires If the wires are longs can act like an antenna and introduce the electrical noise to the unit therefore Do not install the signal input wires in the same conduit with power lines hea...

Page 5: ...ERENCE KEY All models have a label on the rear part with their references printed on the following format LDP 1X4 XX X X Number of digits 4 4 digits Display color R Red G Green Model reference Basic s...

Page 6: ...are made using one screw clamp connector of four terminals located on the rear part of the unit The recommended wire cross section for RTD sensors is 1 mm2 For thermocouples must be used the compensat...

Page 7: ...ion The sensor signal is linearized according to IEC 584 IPTS 1968 ANSI MC96 1 DIN 43710 LDP 1X4 E1 E2 Temperature indicator for thermocouple sensor type E NiCr CuNi Provided with automatic Cold junct...

Page 8: ...OPERATINGTEMPERATURE 0 to 50 C 32 to 122 F STORAGETEMPERATURE 20 to 85 C 4 to 185 F RELATIVEHUMIDITY 0 to 85 not condensed PROTECTION IP65 Front part only MECHANICAL DIMENSIONS See table in page 11 PA...

Page 9: ...l FUSE and spare fuse 8 2 PROTECTION FUSES The unit has a protection fuse located on the power supply socket If this fuse must be replaced or changed because the power supply is changed use the time l...

Page 10: ...Decimal Point Selector must be opened 10 2 SIGNAL CONDITIONERS FOR THERMOCOUPLE SENSORS 9 1 MODEL M1 M2 M3 RTD SENSOR PT 100 TEMPERATURE ACCURACY MODEL TEMPERATURE ACCURACY RESOLUTION RANGE RANGE Sig...

Page 11: ...e Thermocouple simulator to the signal input connector See paragraph 8 4 3 Remove the rear cover taking off all the screws to obtain access to the control board See page 7 paragraph 7 4 Power up the i...

Page 12: ...he drawing below 13 MECHANICAL DIMENSIONS mm inches 123456789012345678901234567 123456789012345678901234567 123456789012345678901234567 123456789012345678901234567 123456789012345678901234567 12345678...

Page 13: ...shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold by OMEGA...

Page 14: ...lewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Conductivity Equipment...

Reviews: