6.7.3
Thermocouple Configuration Interface (TC Models Only)
Thermocouple compatible models of the iServer 2 offer a Cold Junction Calibration feature
in the web UI. To begin cold junction calibration, follow the steps below:
Note:
Cold Junction Calibration is not available for thermocouple types: R, S, and B.
Figure 43: Thermocouple Cold Junction Calibration
Step 1:
Click on the channel readings for the input that will undergo cold junction calibration.
Step 2:
Click on the Cold Junction button.
Step 3:
Click the Proceed button.
Step 4:
Once notified by the web UI that the calibration was successful, click OK to finalize the
changes.